Purpose: This study aimed to evaluate the pressure distribution and comfort of transtibial prosthesis wearers using an affordable ethyl-vinyl acetate (EVA) roll-on (AERO) liner.

Method: Fifteen unilateral transtibial prosthesis users wore patella tendon bearing (PTB) sockets with a polyethylene foam (PE-lite) liner were enrolled this study. AERO liners were provided to all participants. Six force sensors were applied to the residual limb to evaluate pressure distribution during treadmill walking, and the socket comfort score (SCS) was used to evaluate comfortability. Fourier transform infrared (FT-IR) spectroscopy was performed on the EVA and PE-lite liners.

Results: Eleven participants used prefabricated AERO liners and four participants used custom-made AERO liners. The pressure distribution was analysed by the coefficient of variation (CV): PE-lite was 75.7 ± 6.0 and AERO liner 83.3 ± 4.1. Residual limb pressure was significantly decreased when using the AERO liner ( = .0007), with a large effect size ( = 0.87). Mean SCS was 7.5 ± 1.3 and 8.9 ± 1.1 for PE-lite and AERO liner respectively.

Conclusion: Better pressure distribution and comfort were observed when the participants used the AERO liner. AERO had a greater proportion of calcium carbonate (CaCO). These findings suggest that the AERO liner is a better off-the-shelf option for persons using traditional prosthetic sockets and liners.

References
1.
Al-Fakih E, Abu Osman N, Mahmad Adikan F . Techniques for Interface Stress Measurements within Prosthetic Sockets of Transtibial Amputees: A Review of the Past 50 Years of Research. Sensors (Basel). 2016; 16(7). PMC: 4970162. DOI: 10.3390/s16071119. View

2.
Safari M, Meier M . Systematic review of effects of current transtibial prosthetic socket designs-Part 1: Qualitative outcomes. J Rehabil Res Dev. 2015; 52(5):491-508. DOI: 10.1682/JRRD.2014.08.0183. View

3.
Sanders J, Cagle J, Allyn K, Harrison D, Ciol M . How do walking, standing, and resting influence transtibial amputee residual limb fluid volume?. J Rehabil Res Dev. 2014; 51(2):201-12. PMC: 4435803. DOI: 10.1682/JRRD.2013.04.0085. View

4.
Safari M, Meier M . Systematic review of effects of current transtibial prosthetic socket designs--Part 2: Quantitative outcomes. J Rehabil Res Dev. 2015; 52(5):509-26. DOI: 10.1682/JRRD.2014.08.0184. View

5.
Wurdeman S, Stevens P, Campbell J . Mobility Analysis of AmpuTees (MAAT 6): Mobility, Satisfaction, and Quality of Life among Long-Term Dysvascular/Diabetic Prosthesis Users-Results of a Cross-Sectional Analysis. J Prosthet Orthot. 2021; 33(3):161-167. PMC: 8216599. DOI: 10.1097/JPO.0000000000000304. View