Stem Alignment With the Tapered Wedge Cementless Stem "Accolade II" Does Not Affect Bone Mineral Density Over 5 Years After Total Hip Arthroplasty
Overview
Affiliations
Background: This study aimed to investigate the relationship between 3-dimensional (3D) alignment and postoperative bone mineral density (BMD) changes with Accolade II tapered wedge stems, which have a different proximal shape from other tapered wedge stems, up to 5 years after primary total hip arthroplasty.
Methods: We retrospectively analyzed the hips of 89 patients who underwent total hip arthroplasty using the second-generation proximally coated cementless stem (Accolade II; Stryker Orthopedics, Mahwah, NJ) at our institution from 2014 to 2018 over a 5-year follow-up. We evaluated the relationship between stem alignment, measured using 3D-templating software, and BMD changes in the 7 Gruen zones and compared the data with those from a former study using other short taper-wedge stems.
Results: BMDs in zones 1 and 7 continued to decrease gradually every year after surgery, and BMD in zone 7 showed the largest decrease (21%) from baseline over 5 years. No correlation was found between stem alignment (varus/valgus, flexion/extension, and anteversion/retroversion) and changes in BMD in each zone over 5 years.
Conclusions: Our data showed no correlation between 3D stem alignment and changes in BMD in each Gruen zone over 5 years. This suggests that the Accolade II stem may fit better into any shape of the proximal medullary canal because of its unique characteristics.
Onoi Y, Hayashi S, Fujishiro T, Kuroda Y, Nakano N, Hiranaka T Arch Orthop Trauma Surg. 2024; 144(8):3857-3864.