» Articles » PMID: 38732834

Large-Dynamic-Range Ocular Aberration Measurement Based on Deep Learning with a Shack-Hartmann Wavefront Sensor

Overview
Journal Sensors (Basel)
Publisher MDPI
Specialty Biotechnology
Date 2024 May 11
PMID 38732834
Authors
Affiliations
Soon will be listed here.
Abstract

The Shack-Hartmann wavefront sensor (SHWFS) is widely utilized for ocular aberration measurement. However, large ocular aberrations caused by individual differences can easily make the spot move out of the range of the corresponding sub-aperture in SHWFS, rendering the traditional centroiding method ineffective. This study applied a novel convolutional neural network (CNN) model to wavefront sensing for large dynamic ocular aberration measurement. The simulation results demonstrate that, compared to the modal method, the dynamic range of our method for main low-order aberrations in ocular system is increased by 1.86 to 43.88 times in variety. Meanwhile, the proposed method also has the best measurement accuracy, and the statistical root mean square (RMS) of the residual wavefronts is 0.0082 ± 0.0185 λ (mean ± standard deviation). The proposed method generally has a higher accuracy while having a similar or even better dynamic range as compared to traditional large-dynamic schemes. On the other hand, compared with recently developed deep learning methods, the proposed method has a much larger dynamic range and better measurement accuracy.

Citing Articles

Measurement of ocular aberration in noise based on deep learning with a Shack-Hartmann wavefront sensor.

Zhang H, Yang Y, Zhang Z, Yin C, Wang S, Wei K Biomed Opt Express. 2024; 15(11):6531-6548.

PMID: 39553873 PMC: 11563320. DOI: 10.1364/BOE.541483.

References
1.
Guo H, Korablinova N, Ren Q, Bille J . Wavefront reconstruction with artificial neural networks. Opt Express. 2009; 14(14):6456-62. DOI: 10.1364/oe.14.006456. View

2.
Smith D, Greivenkamp J . Generalized method for sorting Shack-Hartmann spot patterns using local similarity. Appl Opt. 2008; 47(25):4548-54. DOI: 10.1364/ao.47.004548. View

3.
Yoon G, Pantanelli S, Nagy L . Large-dynamic-range Shack-Hartmann wavefront sensor for highly aberrated eyes. J Biomed Opt. 2006; 11(3):30502. DOI: 10.1117/1.2197860. View

4.
Thibos L, Bradley A, Hong X . A statistical model of the aberration structure of normal, well-corrected eyes. Ophthalmic Physiol Opt. 2002; 22(5):427-33. DOI: 10.1046/j.1475-1313.2002.00059.x. View

5.
Roggemann M, Schulz T . Algorithm to increase the largest aberration that can be reconstructed from Hartmann sensor measurements. Appl Opt. 2008; 37(20):4321-9. DOI: 10.1364/ao.37.004321. View