Real-world Validation of Smartphone-based Photoplethysmography for Rate and Rhythm Monitoring in Atrial Fibrillation

Abstract

Aims: Photoplethysmography- (PPG) based smartphone applications facilitate heart rate and rhythm monitoring in patients with paroxysmal and persistent atrial fibrillation (AF). Despite an endorsement from the European Heart Rhythm Association, validation studies in this setting are lacking. Therefore, we evaluated the accuracy of PPG-derived heart rate and rhythm classification in subjects with an established diagnosis of AF in unsupervised real-world conditions.

Methods And Results: Fifty consecutive patients were enrolled, 4 weeks before undergoing AF ablation. Patients used a handheld single-lead electrocardiography (ECG) device and a fingertip PPG smartphone application to record 3907 heart rhythm measurements twice daily during 8 weeks. The ECG was performed immediately before and after each PPG recording and was given a diagnosis by the majority of three blinded cardiologists. A consistent ECG diagnosis was exhibited along with PPG data of sufficient quality in 3407 measurements. A single measurement exhibited good quality more often with ECG (93.2%) compared to PPG (89.5%; P < 0.001). However, PPG signal quality improved to 96.6% with repeated measurements. Photoplethysmography-based detection of AF demonstrated excellent sensitivity [98.3%; confidence interval (CI): 96.7-99.9%], specificity (99.9%; CI: 99.8-100.0%), positive predictive value (99.6%; CI: 99.1-100.0%), and negative predictive value (99.6%; CI: 99.0-100.0%). Photoplethysmography underestimated the heart rate in AF with 6.6 b.p.m. (95% CI: 5.8 b.p.m. to 7.4 b.p.m.). Bland-Altman analysis revealed increased underestimation in high heart rates. The root mean square error was 11.8 b.p.m.

Conclusion: Smartphone applications using PPG can be used to monitor patients with AF in unsupervised real-world conditions. The accuracy of AF detection algorithms in this setting is excellent, but PPG-derived heart rate may tend to underestimate higher heart rates.

Citing Articles

Initial experience with a virtual atrial fibrillation clinic after pulmonary vein isolation using follow-up with photoplethysmography.

Reijrink-de Boer M, Wolsink I, Frenaij I, Beukema K, Brouns B, van Dijk V Neth Heart J. 2025; 33(3):85-92.

PMID: 39932623 PMC: 11845631. DOI: 10.1007/s12471-025-01935-6.


Evidence Gaps and Lessons in the Early Detection of Atrial Fibrillation: A Prospective Study in a Primary Care Setting (PREFATE Study).

Clua-Espuny J, Hernandez-Pinilla A, Gentille-Lorente D, Muria-Subirats E, Forcadell-Arenas T, de Diego-Cabanes C Biomedicines. 2025; 13(1).

PMID: 39857703 PMC: 11759169. DOI: 10.3390/biomedicines13010119.


Improving atrial fibrillation or flutter detection and management by smartphone-based photoplethysmography rhythm monitoring following cardiac surgery: a pragmatic randomized trial.

Gruwez H, De Melio N, Vermunicht P, van Langenhoven L, Desteghe L, Lamberigts M Europace. 2025; 27(2).

PMID: 39823508 PMC: 11826921. DOI: 10.1093/europace/euaf015.


Comparative Evaluation of Consumer Wearable Devices for Atrial Fibrillation Detection: Validation Study.

Wouters F, Gruwez H, Smeets C, Pijalovic A, Wilms W, Vranken J JMIR Form Res. 2025; 9():e65139.

PMID: 39791483 PMC: 11737281. DOI: 10.2196/65139.


Smartphone app-based approximation of time spent with atrial fibrillation and symptoms in patients after catheter ablation: data from the TeleCheck-AF project.

Sandgren E, Hermans A, Gawalko M, Betz K, Sohaib A, Fung C Europace. 2024; 26(10).

PMID: 39344253 PMC: 11481324. DOI: 10.1093/europace/euae247.


References
1.
Perez M, Mahaffey K, Hedlin H, Rumsfeld J, Garcia A, Ferris T . Large-Scale Assessment of a Smartwatch to Identify Atrial Fibrillation. N Engl J Med. 2019; 381(20):1909-1917. PMC: 8112605. DOI: 10.1056/NEJMoa1901183. View

2.
Wormann J, Schipper J, Luker J, van den Bruck J, Filipovic K, Erlhofer S . Comparison of pulsed-field ablation versus very high power short duration-ablation for pulmonary vein isolation. J Cardiovasc Electrophysiol. 2023; 34(12):2417-2424. DOI: 10.1111/jce.16101. View

3.
Hermans A, Isaksen J, Gawalko M, Pluymaekers N, van der Velden R, Snippe H . Accuracy of continuous photoplethysmography-based 1 min mean heart rate assessment during atrial fibrillation. Europace. 2023; 25(3):835-844. PMC: 10062358. DOI: 10.1093/europace/euad011. View

4.
Hindricks G, Potpara T, Dagres N, Arbelo E, Bax J, Blomstrom-Lundqvist C . 2020 ESC Guidelines for the diagnosis and management of atrial fibrillation developed in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS): The Task Force for the diagnosis and management of atrial fibrillation of the.... Eur Heart J. 2020; 42(5):373-498. DOI: 10.1093/eurheartj/ehaa612. View

5.
Marcus G . Smartwatch-Detected Atrial Fibrillation: The "Value" in the Positive Predictive Value. Circulation. 2022; 146(23):1733-1734. DOI: 10.1161/CIRCULATIONAHA.122.062292. View