» Articles » PMID: 38591559

Comparing Torque Transmission of Different Bracket Systems in Combination with Various Archwires Considering Play in the Bracket Slot: An In Vitro Study

Overview
Publisher MDPI
Date 2024 Apr 9
PMID 38591559
Authors
Affiliations
Soon will be listed here.
Abstract

This study aims to examine the play between various archwires and bracket systems, exploring potential variations in angle values for specific torque and torque values for a given angle along different bracket systems. Therefore, seven brackets systems were evaluated in conjunction with different stainless steel archwires of varying dimensions (0.016″ × 0.022″, 0.018″ × 0.025″, and 0.019″ × 0.025″). Biomechanical behavior during torque development and transmission was assessed using a six-component force/torque sensor. Torque angles (5-45°) were specified with subsequent torque measurement, and the sequence was reversed by setting the torque (5-30 Nmm) and measuring the angle. A reference measurement with 0 Nmm torque served to evaluate bracket slot play. Bracket play (0 Nmm) during palatal load ranged between 20.06° and 32.50° for 0.016″ × 0.022″ wire, 12.83° and 21.11° for 0.018″ × 0.025″ wire, and 8.39° and 18.73° for 0.019″ × 0.025″ wire. The BioQuick bracket exhibited the highest play, while Wave SL and Damon Q brackets demonstrated the lowest play ( < 0.001). Significant differences ( < 0.001) between the brackets were observed in the torque angles required to achieve torques of 5-20 Nmm. In summary, each bracket system has a different torque transmission, which is of great clinical importance in order to achieve correct torque transmission and avoid complications such as root resorption.

References
1.
Hirai M, Nakajima A, Kawai N, Tanaka E, Igarashi Y, Sakaguchi M . Measurements of the torque moment in various archwire-bracket-ligation combinations. Eur J Orthod. 2011; 34(3):374-80. DOI: 10.1093/ejo/cjr022. View

2.
Al-Thomali Y, Mohamed R, Basha S . Torque expression in self-ligating orthodontic brackets and conventionally ligated brackets: A systematic review. J Clin Exp Dent. 2017; 9(1):e123-e128. PMC: 5268106. DOI: 10.4317/jced.53187. View

3.
Huang Y, Keilig L, Rahimi A, Reimann S, Eliades T, Jager A . Numeric modeling of torque capabilities of self-ligating and conventional brackets. Am J Orthod Dentofacial Orthop. 2009; 136(5):638-43. DOI: 10.1016/j.ajodo.2009.04.018. View

4.
Sifakakis I, Pandis N, Makou M, Eliades T, Katsaros C, Bourauel C . Torque efficiency of different archwires in 0.018- and 0.022-inch conventional brackets. Angle Orthod. 2013; 84(1):149-54. PMC: 8683062. DOI: 10.2319/021713-138.1. View

5.
Bartley N, Turk T, Colak C, Elekdag-Turk S, Jones A, Petocz P . Physical properties of root cementum: Part 17. Root resorption after the application of 2.5° and 15° of buccal root torque for 4 weeks: a microcomputed tomography study. Am J Orthod Dentofacial Orthop. 2011; 139(4):e353-60. DOI: 10.1016/j.ajodo.2010.01.033. View