Improvement of Reproducibility of Magnetic Moment Detected by a SQUID Magnetometer Through Radial Offset Measurement on a YIG Sphere

Overview
Journal IEEE Trans Magn
Date 2024 Mar 15
PMID 38486836
Authors
Affiliations
Soon will be listed here.
Abstract

We report here on the reproducibility of measurements on a second-order gradiometer superconducting quantum interference device magnetometer of two different yttrium iron garnet spheres, both having a diameter of 1 mm: 1) the National Institute of Standards and Technology magnetic moment standard reference material (SRM) and 2) a commercial sample. It has been suggested that rotating the sample rod around its axis can move the sample center toward the center of the second-order gradiometer coil. The observed value of the magnetic moment will be theoretically a minimum when the radial offset is 0, and this value will increase in a "quadratic" manner with the radial offset. When the magnetic moment of the SRM was repeatedly measured as a function of rotation angle from 0° to 360°, we observed a sinusoidal variation in the measured values. The radial offset dependence of the observed magnetic moment was experimentally confirmed by the measurements of the commercial sphere placed in a hole in several cylindrical containers, wherein the distance between the center of the hole and the center of the container was . The -dependence of the minimum from each -dependent measurement series is qualitatively consistent with the theoretical curve. When the -dependent measurements for the SRM in a capsule were repeated 12× over 21 months, the relative standard deviation of the minimums improved up to 0.1%. Knowledge of these facts will be necessary for the accurate measurement of the magnetic moment of other sample forms (e.g., powders).

References
1.
Boekelheide Z, Dennis C . Artifacts in magnetic measurements of fluid samples. AIP Adv. 2017; 6(8). PMC: 5531296. DOI: 10.1063/1.4960457. View

2.
Matsumoto N, Shimosaka T . Low-temperature Electronic Paramagnetic Resonance Measurements of TEMPO and 4-Hydroxy-TEMPO Benzoate for Purity Analyses by the Effective Magnetic-moment Method. Anal Sci. 2017; 33(9):1059-1065. DOI: 10.2116/analsci.33.1059. View

3.
Matsumoto N, Itoh N . Measuring Number of Free Radicals and Evaluating the Purity of Di(phenyl)-(2,4,6-trinitrophenyl)iminoazanium [DPPH] Reagents by Effective Magnetic Moment Method. Anal Sci. 2018; 34(8):965-971. DOI: 10.2116/analsci.18P120. View