So, What's Best? Accuracy and Acceptance of Thermometers in Triage and Inpatients in a Low-resource Tropical Setting - The MaTe Study

Overview
Journal Heliyon
Specialty Social Sciences
Date 2024 Feb 19
PMID 38371989
Authors
Affiliations
Soon will be listed here.
Abstract

Objectives: We searched for the most-suitable thermometry method in the low-resource, tropical setting of Sierra Leone, both in terms of accuracy and also patient and user acceptance.

Methods: We conducted a prospective comparative study of different methods of body temperature measurement. Each participant had their temperature taken by four different methods: non-contact infrared temperature (NCIT), axillary, tympanic membrane and rectal measurements. Rectal temperature was considered clinical gold standard. Primary outcome was predicted sensitivity and specificity of thermometry methods in detecting fever (rectal temperature ≥38.0 °C). Questionnaires were used to explore patient and healthcare worker attitudes towards different methods of temperature-taking.

Results: 824 rectal body temperature readings were taken from 562 participants. The mean rectal temperature was 37.4 °C (IQR 37 °C to 37.7 °C), with a minimum reading of 35.2 °C and maximum of 41.0 °C. Tympanic membrane thermometry showed the highest sensitivity of fever detection using the Genius3 TM thermometer (sensitivity 70.8 %, 95 % CI 60.2%-79.9 %; specificity 97.2 %, 95 % CI 95.5-98.4 %); and Braun TM (sensitivity 51.5 %, 95 % CI 42.6%-62.0 %; specificity 98.8 %, 95 % CI 97.7-99.5). NCIT thermometry sensitivity was low (36.8 %-41.4 % for the two devices used). Axillary thermometry sensitivity was 40.6 %. Participants ranked NCIT as the most and rectal as the least preferred method. Questionnaires from 32 participating nurses showed agreeability to using NCIT, TM and axillary methods routinely, but less so for rectal thermometry.

Conclusions: When combining the accuracy of different thermometry methods in detecting fever with user and patient acceptability, tympanic membrane thermometry appears most suitable but also has limitations.

Citing Articles

Diagnostic Accuracy of Three Methods of Body Temperature Measurement in Children: A Latent Class Approach.

Swapna K, George B, Pillai R, Rajasekharan Nair J Cureus. 2025; 17(1):e77325.

PMID: 39935916 PMC: 11812845. DOI: 10.7759/cureus.77325.

References
1.
. Global, regional, and national age-sex-specific mortality for 282 causes of death in 195 countries and territories, 1980-2017: a systematic analysis for the Global Burden of Disease Study 2017. Lancet. 2018; 392(10159):1736-1788. PMC: 6227606. DOI: 10.1016/S0140-6736(18)32203-7. View

2.
Kelechi O, Benedict E, Nwolisa E, Ifeyinwa A, Seline O . Acceptance of temporal artery thermometry by Nigerian mothers: a comparison with the traditional methods. Pan Afr Med J. 2015; 19:263. PMC: 4382067. DOI: 10.11604/pamj.2014.19.263.2899. View

3.
van Staaij B, Rovers M, Schilder A, Hoes A . Accuracy and feasibility of daily infrared tympanic membrane temperature measurements in the identification of fever in children. Int J Pediatr Otorhinolaryngol. 2003; 67(10):1091-7. DOI: 10.1016/s0165-5876(03)00198-8. View

4.
Donkers L, van Furth A, van der Zwet W, Fetter W, Roord J, Vandenbroucke-Grauls C . [Enterobacter cloacae epidemic on a neonatal intensive care unit due to the use of contaminated thermometers]. Ned Tijdschr Geneeskd. 2001; 145(13):643-7. View

5.
Barnett B, Nunberg S, Tai J, Lesser M, Fridman V, Nichols P . Oral and tympanic membrane temperatures are inaccurate to identify Fever in emergency department adults. West J Emerg Med. 2012; 12(4):505-11. PMC: 3236177. DOI: 10.5811/westjem.2011.2.1963. View