» Articles » PMID: 38361025

Synthetic Interpolated DSA for Radiation Exposure Reduction Via Gamma Variate Contrast Flow Modeling: a Retrospective Cohort Study

Overview
Journal Eur Radiol Exp
Date 2024 Feb 16
PMID 38361025
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Digital subtraction angiography (DSA) yields high cumulative radiation dosages (RD) delivered to patients. We present a temporal interpolation of low frame rate angiograms as a method to reduce cumulative RDs.

Methods: Patients undergoing interventional evaluation and treatment of cerebrovascular vasospasm following subarachnoid hemorrhage were retrospectively identified. DSAs containing pre- and post-intervention runs capturing the full arterial, capillary, and venous phases with at least 16 frames each were selected. Frame rate reduction (FRR) of the original DSAs was performed to 50%, 66%, and 75% of the original frame rate. Missing frames were regenerated by sampling a gamma variate model (GVM) fit to the contrast response curves to the reduced data. A formal reader study was performed to assess the diagnostic accuracy of the "synthetic" studies (sDSA) compared to the original DSA.

Results: Thirty-eight studies met inclusion criteria (average RD 1,361.9 mGy). Seven were excluded for differing views, magnifications, or motion. GVMs fit to 50%, 66%, and 75% FRR studies demonstrated average voxel errors of 2.0 ± 2.5% (mean ± standard deviation), 6.5 ± 1.5%, and 27 ± 2%, respectively for anteroposterior projections, 2.0 ± 2.2%, 15.0 ± 3.1%, and 14.8 ± 13.0% for lateral projections, respectively. Reconstructions took 0.51 s/study. Reader studies demonstrated an average rating of 12.8 (95% CI 12.3-13.3) for 75% FRR, 12.7 (12.2-13.2) for 66% FRR and 12.0 (11.5-12.5) for 50% FRR using Subjective Image Grading Scale. Kendall's coefficient of concordance resulted in W = 0.506.

Conclusion: FRR by 75% combined with GVM reconstruction does not compromise diagnostic quality for the assessment of cerebral vasculature.

Relevance Statement: Using this novel algorithm, it is possible to reduce the frame rate of DSA by as much as 75%, with a proportional reduction in radiation exposure, without degrading imaging quality.

Key Points: • DSA delivers some of the highest doses of radiation to patients. • Frame rate reduction (FRR) was combined with bolus tracking to interpolate intermediate frames. • This technique provided a 75% FRR with preservation of diagnostic utility as graded by a formal reader study for cerebral angiography performed for the evaluation of cerebral vasospasm. • This approach can be applied to other types of angiography studies.

References
1.
Kahn E, Gemmete J, Chaudhary N, Thompson B, Chen K, Christodoulou E . Radiation dose reduction during neurointerventional procedures by modification of default settings on biplane angiography equipment. J Neurointerv Surg. 2015; 8(8):819-23. DOI: 10.1136/neurintsurg-2015-011891. View

2.
Pearl M, Torok C, Wang J, Wyse E, Mahesh M, Gailloud P . Practical techniques for reducing radiation exposure during cerebral angiography procedures. J Neurointerv Surg. 2014; 7(2):141-5. DOI: 10.1136/neurintsurg-2013-010982. View

3.
Scalzo F, Liebeskind D . Perfusion Angiography in Acute Ischemic Stroke. Comput Math Methods Med. 2016; 2016:2478324. PMC: 4947508. DOI: 10.1155/2016/2478324. View

4.
Axel L . Cerebral blood flow determination by rapid-sequence computed tomography: theoretical analysis. Radiology. 1980; 137(3):679-86. DOI: 10.1148/radiology.137.3.7003648. View

5.
Johnson G, Wetzel S, Cha S, Babb J, Tofts P . Measuring blood volume and vascular transfer constant from dynamic, T(2)*-weighted contrast-enhanced MRI. Magn Reson Med. 2004; 51(5):961-8. DOI: 10.1002/mrm.20049. View