A Modified Technique for Transcatheter Pulmonary Valve Implantation of SAPIEN 3 Valves in Large Right Ventricular Outflow Tract: A Matched Comparison Study

Overview
Journal J Clin Med
Specialty General Medicine
Date 2023 Dec 23
PMID 38137725
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: Percutaneous pulmonary valve implantation (PPVI) with a SAPIEN 3 valve is effective for treating treat right ventricle outflow (RVOT) dysfunction. A modified technique was developed without prestenting using a protective valve delivery method. We aimed to compare the procedural results of the modified technique group (MTG) to those of patients in a conventional technique group (CTG).

Methods: We designed a matched before-after study. All consecutive PPVI with SAPIEN 3 performed in the MTG over 9 months were matched, based on the RVOT type and size, to consecutive procedures performed previously with SAPIEN 3.

Results: A total of 54 patients were included, equally distributed in the two groups. The sizes of the SAPIEN 3 valves were 23 mm (n = 9), 26 mm (n = 9), 29 mm (n = 36). The two groups were similar regarding demographic data, RVOT type, and pre-procedure hemodynamics. PPVI was performed in a single procedure in all patients of the MTG, whereas six (22.2%) patients of the CTG group underwent prestenting as a first step and valve implantation later ( = 0.02). The procedures were successful in all cases. Stent embolization was reported in two patients (7.4%) in the CTG, which were impacted in pulmonary arteries. In one case (3.7%), in the MTG, an unstable 29 mm SAPIEN 3 valve was stabilized with two stents and additional valve-in-valve implantation. The hemodynamics results were good in all cases, without significant differences between the two groups. The procedures' durations and fluoroscopy times were significantly reduced in the MTG (48.1 versus 82.6 min, < 0.0001; 15.2 versus 29.8 min, = 0.0002). During follow-up, neither stent fracture nor valve dysfunction was noticed in either group.

Conclusion: PPVI without prestenting and with a protective delivery method of the SAPIEN 3 valve significantly reduces the procedure's complexity, the duration, and the irradiation while maintaining excellent hemodynamics results in selected cases.

Citing Articles

Standardized Bench Test Evaluation of Biomechanical Characteristics of Stents Used in Right Ventricular Outflow Tract Revalvulation.

Piriou P, Plessis J, Manigold T, Letocart V, Le Ruz R, Padovani P Cardiovasc Eng Technol. 2024; 15(4):443-450.

PMID: 38468115 DOI: 10.1007/s13239-024-00726-1.

References
1.
Husain J, Praichasilchai P, Gilbert Y, Qureshi S, Morgan G . Early European experience with the Venus P-valve®: filling the gap in percutaneous pulmonary valve implantation. EuroIntervention. 2016; 12(5):e643-51. DOI: 10.4244/EIJV12I5A105. View

2.
Hascoet S, Dalla Pozza R, Bentham J, Carere R, Kanaan M, Ewert P . Early outcomes of percutaneous pulmonary valve implantation using the Edwards SAPIEN 3 transcatheter heart valve system. EuroIntervention. 2018; 14(13):1378-1385. DOI: 10.4244/EIJ-D-18-01035. View

3.
Morgan G, Sadeghi S, Salem M, Wilson N, Kay J, Rothman A . SAPIEN valve for percutaneous transcatheter pulmonary valve replacement without "pre-stenting": A multi-institutional experience. Catheter Cardiovasc Interv. 2018; 93(2):324-329. DOI: 10.1002/ccd.27932. View

4.
Faccini A, Butera G . Tricuspid regurgitation as a complication of Edwards Sapien XT valve implantation in pulmonary position a problem to deal with. Catheter Cardiovasc Interv. 2018; 91(5):927-931. DOI: 10.1002/ccd.27527. View

5.
Ghobrial J, Levi D, Aboulhosn J . Native Right Ventricular Outflow Tract Transcatheter Pulmonary Valve Replacement Without Pre-Stenting. JACC Cardiovasc Interv. 2018; 11(6):e41-e44. DOI: 10.1016/j.jcin.2017.11.014. View