Assessment of Measurement Accuracy of Amplified DNA Using a Colorimetric Loop-mediated Isothermal Amplification Assay
Overview
Biotechnology
Affiliations
A colorimetric loop-mediated isothermal amplification assay detects changes in pH during amplification based on color changes at a constant temperature. Currently, various studies have focused on developing and assessing molecular point-of-care testing instruments. In this study, we evaluated amplified DNA concentrations measured using the colorimetric LAMP assay of the 1POT™ Professional device (1drop Inc, Korea). Results of the 1POT analysis of clinical samples were compared with measurements obtained from the Qubit™ 4 and NanoDrop™ 2000 devices (both from Thermo Fisher Scientific, MA, USA). These results showed a correlation of 0.98 (95% CI: 0.96-0.99) and 0.96 (95% CI: 0.92-0.98) between 1POT and the Qubit and NanoDrop. 1POT can measure amplified DNA accurately and is suitable for on-site molecular diagnostics.
An accurate and convenient method for via one-step LAMP-CRISPR/Cas12b detection platform.
Liu T, Liu Q, Chen F, Shi Y, Maimaiti G, Yang Z Front Cell Infect Microbiol. 2024; 14:1409078.
PMID: 39176261 PMC: 11338869. DOI: 10.3389/fcimb.2024.1409078.