» Articles » PMID: 38010083

Optimization of the Flip Angles of Narrow-band Editing Pulses in J-difference Edited MRS of Lactate at 3T

Overview
Journal Magn Reson Med
Publisher Wiley
Specialty Radiology
Date 2023 Nov 27
PMID 38010083
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Application of highly selective editing RF pulses provides a means of minimizing co-editing of contaminants in J-difference MRS (MEGA), but it causes reduction in editing yield. We examined the flip angles (FAs) of narrow-band editing pulses to maximize the lactate edited signal with minimal co-editing of threonine.

Methods: The effect of editing-pulse FA on the editing performance was examined, with numerical and phantom analyses, for bandwidths of 17.6-300 Hz in MEGA-PRESS editing of lactate at 3T. The FA and envelope of 46 ms Gaussian editing pulses were tailored to maximize the lactate edited signal at 1.3 ppm and minimize co-editing of threonine. The optimized editing-pulse FA MEGA scheme was tested in brain tumor patients.

Results: Simulation and phantom data indicated that the optimum FA of MEGA editing pulses is progressively larger than 180° as the editing-pulse bandwidth decreases. For 46 ms long 17.6 Hz bandwidth Gaussian pulses and other given sequence parameters, the lactate edited signal was maximum at the first and second editing-pulse FAs of 241° and 249°, respectively. The edit-on and difference-edited lactate peak areas of the optimized FA MEGA were greater by 43% and 25% compared to the 180°-FA MEGA, respectively. In-vivo data confirmed the simulation and phantom results. The lesions of the brain tumor patients showed elevated lactate and physiological levels of threonine.

Conclusion: The lactate MEGA editing yield is significantly increased with editing-pulse FA much larger than 180° when the editing-pulse bandwidth is comparable to the lactate quartet frequency width.

References
1.
Askari P, Dimitrov I, Ganji S, Tiwari V, Levy M, Patel T . Spectral fitting strategy to overcome the overlap between 2-hydroxyglutarate and lipid resonances at 2.25 ppm. Magn Reson Med. 2021; 86(4):1818-1828. PMC: 8295210. DOI: 10.1002/mrm.28829. View

2.
Ganji S, An Z, Tiwari V, Chang Y, Patel T, Maher E . Optimization of spectrally selective 180° radiofrequency pulse timings in J-difference editing (MEGA) of lactate. Magn Reson Med. 2021; 87(3):1150-1164. PMC: 8776585. DOI: 10.1002/mrm.29051. View

3.
Ogg R, Kingsley P, Taylor J . WET, a T1- and B1-insensitive water-suppression method for in vivo localized 1H NMR spectroscopy. J Magn Reson B. 1994; 104(1):1-10. DOI: 10.1006/jmrb.1994.1048. View

4.
Choi C, Ganji S, Madan A, Hulsey K, An Z, Zhang S . In vivo detection of citrate in brain tumors by 1H magnetic resonance spectroscopy at 3T. Magn Reson Med. 2013; 72(2):316-23. PMC: 4236907. DOI: 10.1002/mrm.24946. View

5.
Cai M, Wang H, Song H, Yang R, Wang L, Xue X . Lactate Is Answerable for Brain Function and Treating Brain Diseases: Energy Substrates and Signal Molecule. Front Nutr. 2022; 9:800901. PMC: 9099001. DOI: 10.3389/fnut.2022.800901. View