Large Articulations Do Not Increase Wear Rates of Thin Second-generation Highly Cross-linked Polyethylene Liners at Ten Years

Overview
Journal Bone Jt Open
Date 2023 Nov 5
PMID 37926113
Authors
Affiliations
Soon will be listed here.
Abstract

Aims: Radiostereometric analysis (RSA) is the most accurate radiological method to measure in vivo wear of highly cross-linked polyethylene (XLPE) acetabular components. We have previously reported very low wear rates for a sequentially irradiated and annealed X3 XLPE liner (Stryker Orthopaedics, USA) when used in conjunction with a 32 mm femoral heads at ten-year follow-up. Only two studies have reported the long-term wear rate of X3 liners used in conjunction with larger heads using plain radiographs which have poor sensitivity. The aim of this study was to measure the ten-year wear of thin X3 XLPE liners against larger 36 or 40 mm articulations with RSA.

Methods: We prospectively reviewed 19 patients who underwent primary cementless THA with the XLPE acetabular liner (X3) and a 36 or 40 mm femoral head with a resultant liner thickness of at least 5.8 mm. RSA radiographs at one week, six months, and one, two, five, and ten years postoperatively and femoral head penetration within the acetabular component were measured with UmRSA software. Of the initial 19 patients, 12 were available at the ten-year time point.

Results: The median proximal, 2D, and 3D wear rates calculated between one and ten years were all less than 0.005 mm/year, with no patient recording a proximal wear rate of more than 0.021 mm/year. Importantly, there was no increase in the wear rate between five and ten years.

Conclusion: The very low wear rate of X3 XLPE liners with larger articulations remains encouraging for the future clinical performance of this material.

Citing Articles

Safety of using a large femoral head on thin polyethylene for total hip arthroplasty based on different types of polyethylene.

Do M, Moon N, Suh K, Kim J, Lee S, Shin W Sci Rep. 2023; 13(1):22769.

PMID: 38123795 PMC: 10733297. DOI: 10.1038/s41598-023-50217-x.

References
1.
Selvarajah E, Hooper G, Grabowski K, Frampton C, Woodfield T, Inglis G . The rates of wear of X3 highly cross-linked polyethylene at five years when coupled with a 36 mm diameter ceramic femoral head in young patients. Bone Joint J. 2015; 97-B(11):1470-4. DOI: 10.1302/0301-620X.97B11.35736. View

2.
Teeter M, Yuan X, Somerville L, MacDonald S, McCalden R, Naudie D . Thirteen-year wear rate comparison of highly crosslinked and conventional polyethylene in total hip arthroplasty: long-term follow-up of a prospective randomized controlled trial. Can J Surg. 2017; 60(3):212-216. PMC: 5453765. DOI: 10.1503/cjs.005216. View

3.
Dumbleton J, Manley M, Edidin A . A literature review of the association between wear rate and osteolysis in total hip arthroplasty. J Arthroplasty. 2002; 17(5):649-61. DOI: 10.1054/arth.2002.33664. View

4.
Ito H, Tanino H, Yamanaka Y, Minami A, Matsuno T . Intermediate- to long-term results after hybrid total hip arthroplasty in patients with rheumatoid arthritis. J Arthroplasty. 2012; 28(2):309-14. DOI: 10.1016/j.arth.2012.06.002. View

5.
Callary S, Field J, Campbell D . The rate of wear of second-generation highly crosslinked polyethylene liners five years post-operatively does not increase if large femoral heads are used. Bone Joint J. 2016; 98-B(12):1604-1610. DOI: 10.1302/0301-620X.98B12.37682. View