Significance: Fluorescence guided surgery (FGS) has demonstrated improvements in decision making and patient outcomes for a wide range of surgical procedures. Not only can FGS systems provide a higher level of structural perfusion accuracy in tissue reconstruction cases but they can also serve for real-time functional characterization. Multiple FGS devices have been Food and Drug administration (FDA) cleared for use in open and laparoscopic surgery. Despite the rapid growth of the field, there has been a lack standardization methods.

Aim: This work overviews commonalities inherent to optical imaging methods that can be exploited to produce such a standardization procedure. Furthermore, a system evaluation pipeline is proposed and executed through the use of photo-stable indocyanine green fluorescence phantoms. Five different FDA-approved open-field FGS systems are used and evaluated with the proposed method.

Approach: The proposed pipeline encompasses the following characterization: (1) imaging spatial resolution and sharpness, (2) sensitivity and linearity, (3) imaging depth into tissue, (4) imaging system DOF, (5) uniformity of illumination, (6) spatial distortion, (7) signal to background ratio, (8) excitation bands, and (9) illumination wavelength and power.

Results: The results highlight how such a standardization approach can be successfully implemented for inter-system comparisons as well as how to better understand essential features within each FGS setup.

Conclusions: Despite clinical use being the end goal, a robust yet simple standardization pipeline before clinical trials, such as the one presented herein, should benefit regulatory agencies, manufacturers, and end-users to better assess basic performance and improvements to be made in next generation FGS systems.

Citing Articles

Combined dual-channel fluorescence depth sensing of indocyanine green and protoporphyrin IX kinetics in subcutaneous murine tumors.

Kulkarni M, Reed M, Cao X, Garcia H, Ochoa M, Jiang S J Biomed Opt. 2024; 30(Suppl 1):S13709.

PMID: 39559531 PMC: 11571966. DOI: 10.1117/1.JBO.30.S1.S13709.


Impact of signal-to-noise ratio and contrast definition on the sensitivity assessment and benchmarking of fluorescence molecular imaging systems.

Kriukova E, LaRochelle E, Pfefer T, Kanniyappan U, Gioux S, Pogue B J Biomed Opt. 2024; 30(Suppl 1):S13703.

PMID: 39034959 PMC: 11256003. DOI: 10.1117/1.JBO.30.S1.S13703.


Semi-automatic standardized analysis method to objectively evaluate near-infrared fluorescent dyes in image-guided surgery.

Dijkhuis T, Bijlstra O, Warmerdam M, Faber R, Linders D, Galema H J Biomed Opt. 2024; 29(2):026001.

PMID: 38312853 PMC: 10833575. DOI: 10.1117/1.JBO.29.2.026001.


A Wearable Fluorescence Imaging Device for Intraoperative Identification of Human Brain Tumors.

Mohtasebi M, Huang C, Zhao M, Mazdeyasna S, Liu X, Haratbar S IEEE J Transl Eng Health Med. 2024; 12:225-232.

PMID: 38196823 PMC: 10776094. DOI: 10.1109/JTEHM.2023.3338564.

References
1.
Troyan S, Kianzad V, Gibbs-Strauss S, Gioux S, Matsui A, Oketokoun R . The FLARE intraoperative near-infrared fluorescence imaging system: a first-in-human clinical trial in breast cancer sentinel lymph node mapping. Ann Surg Oncol. 2009; 16(10):2943-52. PMC: 2772055. DOI: 10.1245/s10434-009-0594-2. View

2.
Zhu B, Rasmussen J, Sevick-Muraca E . A matter of collection and detection for intraoperative and noninvasive near-infrared fluorescence molecular imaging: to see or not to see?. Med Phys. 2014; 41(2):022105. PMC: 3987664. DOI: 10.1118/1.4862514. View

3.
Tanyi J, Randall L, Chambers S, Butler K, Winer I, Langstraat C . A Phase III Study of Pafolacianine Injection (OTL38) for Intraoperative Imaging of Folate Receptor-Positive Ovarian Cancer (Study 006). J Clin Oncol. 2022; 41(2):276-284. DOI: 10.1200/JCO.22.00291. View

4.
Suero Molina E, Stogbauer L, Jeibmann A, Warneke N, Stummer W . Validating a new generation filter system for visualizing 5-ALA-induced PpIX fluorescence in malignant glioma surgery: a proof of principle study. Acta Neurochir (Wien). 2020; 162(4):785-793. PMC: 7066295. DOI: 10.1007/s00701-020-04227-7. View

5.
Smith J, Aguenounon E, Gioux S, Intes X . Macroscopic fluorescence lifetime topography enhanced via spatial frequency domain imaging. Opt Lett. 2020; 45(15):4232-4235. PMC: 7935427. DOI: 10.1364/OL.397605. View