Conceptual Design of a Portable, Solid-Nitrogen-Cooled 0.5-T/560-mm Point-of-Care MRI Magnet

Overview
Date 2023 Aug 28
PMID 37638131
Authors
Affiliations
Soon will be listed here.
Abstract

We describe the conceptual design of a portable, liquid-helium-free, all-REBCO, 0.5-T/560-mm point-of-care magnetic resonance imaging (MRI) magnet. It is free from an external power supply and a refrigeration system during operation. In our portable MRI magnet, we use a detachable "cryocirculator" that circulates, in a closed circuit, cold working fluid, and most importantly for portability, it can be readily coupled to or decoupled from the magnet, in contrast, a conventional cryocooler is mechanically attached to the magnet. Another unique feature of our system is a volume of solid nitrogen (SN2) in the cold chamber that adds enough thermal mass to the magnet in the 30-36-K operating temperature range, enabling it to maintain its field over a period of, for this system, ≥10 hours, plenty enough for this portable MRI system, uncoupled from its cryocirculator, to perform its mission before it needs recooling.

References
1.
Ling J, Voccio J, Hahn S, Qu T, Bascunan J, Iwasa Y . A persistent-mode 0.5 T solid-nitrogen-cooled MgB2 magnet for MRI. Supercond Sci Technol. 2017; 30(2). PMC: 5613760. DOI: 10.1088/1361-6668/30/2/024011. View

2.
Runge V, Heverhagen J . Advocating the Development of Next-Generation, Advanced-Design Low-Field Magnetic Resonance Systems. Invest Radiol. 2020; 55(12):747-753. DOI: 10.1097/RLI.0000000000000703. View

3.
Mazurek M, Cahn B, Yuen M, Prabhat A, Chavva I, Shah J . Portable, bedside, low-field magnetic resonance imaging for evaluation of intracerebral hemorrhage. Nat Commun. 2021; 12(1):5119. PMC: 8387402. DOI: 10.1038/s41467-021-25441-6. View

4.
Rutt B, Lee D . The impact of field strength on image quality in MRI. J Magn Reson Imaging. 1996; 6(1):57-62. DOI: 10.1002/jmri.1880060111. View

5.
Campbell-Washburn A, Ramasawmy R, Restivo M, Bhattacharya I, Basar B, Herzka D . Opportunities in Interventional and Diagnostic Imaging by Using High-Performance Low-Field-Strength MRI. Radiology. 2019; 293(2):384-393. PMC: 6823617. DOI: 10.1148/radiol.2019190452. View