How Does Pupil Size Affect Lens and Corneal Densitometry Measured by Scheimpflug Tomography?

Overview
Specialty Ophthalmology
Date 2023 Aug 21
PMID 37602578
Authors
Affiliations
Soon will be listed here.
Abstract

Objectives: To investigate the effects of pupil diameter on the evaluation of lens and corneal densitometry measured by Scheimpflug tomography.

Materials And Methods: This cross-sectional and comparative study used the right eyes of 32 participants. Corneal and lenticular optical densitometries, corneal volume, anterior segment volume, and anterior chamber depth measurements were taken with the Scheimpflug imaging system when the pupils were mid-dilated and fully dilated. The results were statistically compared.

Results: The mean lens density was 19.20±3.05 when the pupils were mid-dilated (mean pupil diameter 2.98±0.89 mm) and 23.25±3.88 at full dilation (mean pupil diameter 5.01±0.92 mm) (p<0.001). The mean corneal density was 16.15±0.99 with mid-dilated pupils and 16.38±0.95 with fully dilated pupils (p=0.065). Anterior chamber depth and anterior segment volume measurements increased with larger pupil diameter (p<0.05).

Conclusion: The lens densitometry values increased with an increase in pupil diameter. The corneal density measurements increased minimally but the differences were not statistically significant. This study revealed that lens densitometry was significantly affected by pupil diameter.

References
1.
Ostrin L, Glasser A . The effects of phenylephrine on pupil diameter and accommodation in rhesus monkeys. Invest Ophthalmol Vis Sci. 2003; 45(1):215-21. PMC: 2913435. DOI: 10.1167/iovs.03-0704. View

2.
Kirkwood B, Hendicott P, Read S, Pesudovs K . Repeatability and validity of lens densitometry measured with Scheimpflug imaging. J Cataract Refract Surg. 2009; 35(7):1210-5. DOI: 10.1016/j.jcrs.2009.03.017. View

3.
Bahar A, Pekel G . How Does Light Intensity of the Recording Room Affect the Evaluation of Lens and Corneal Clarity by Scheimpflug Tomography?. Cornea. 2019; 39(2):137-139. DOI: 10.1097/ICO.0000000000002212. View

4.
Durukan I . Evaluation of corneal and lens clarity in unilateral pseudoexfoliation syndrome: a densitometric analysis. Clin Exp Optom. 2018; 101(6):740-746. DOI: 10.1111/cxo.12802. View

5.
Lei H, Yang K, Sun C, Chen C, Huang B, Ng S . Obtained mydriasis in long-term type 2 diabetic patients. J Ocul Pharmacol Ther. 2011; 27(6):599-602. DOI: 10.1089/jop.2011.0090. View