Serotonin Secretion by Blood Platelets: Accuracy of High-performance Liquid Chromatography-electrochemical Technique Compared with the Isotopic Test and Use in a Clinical Laboratory

Overview
Publisher Elsevier
Date 2023 Aug 21
PMID 37601022
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Mild secretion defects are the most frequent and challenging blood platelet disorders to diagnose. Most δ-granule secretion tests lack validation, are not quantitative, or have unreliable response to weak platelet agonists.

Objectives: To compare platelet serotonin secretion by HPLC-electrochemical detection technique (HPLC-ECD) with the reference isotopic test (H-5-HT), evaluating its performance in clinical laboratories.

Methods: The assay validation followed STARD-2015 recommendations. HPLC-ECD measured the nonsecreted serotonin remaining in platelet pellets after aggregation, comparing it with the reference H-5-HT assay. We studied subjects with inherited and aspirin-induced blood platelet disorders and assessed the HPLC-ECD operation for routine clinical diagnosis.

Results: Calibration curves were linear (R = 0.997), with SD for residuals of 3.91% and analytical sensitivity of 5ng/mL. Intra- and interassay imprecision bias ranged between -8.5% and 2.1% and -9% and 3.1%, respectively. Serotonin recovery and stability were >95%, and the variability range of measurements was -5.5% to 4.6%. Statistical differences detected between tests were biologically irrelevant, with bias of 1.48% (SD, 8.43) and CI agreement of -18% to 15%. Both assays distinctly detected platelet secretion induced by 10 μM epinephrine and 4 μmM adenosine diphosphate. However, HPLC-ECD is quantitative and more sensitive to low serotonin content in blood platelets. Reference cutoffs for each agonist were determined in 87 subjects. Initially, the HPLC-ECD requires relatively expensive equipment and trained operators but has remarkably cheap running costs and a turn-around time of 24-36 hours. We have used this diagnostic tool routinely for >8 years.

Conclusion: HPLC-ECD assay for platelet serotonin secretion is highly accurate, has advantages over the reference H-5-HT test, and is suitable as a clinical laboratory technique.

References
1.
Fouassier M, Babuty A, Debord C, Bene M . Platelet immunophenotyping in health and inherited bleeding disorders, a review and practical hints. Cytometry B Clin Cytom. 2020; 98(6):464-475. DOI: 10.1002/cyto.b.21892. View

2.
Lotta L, Maino A, Tuana G, Rossio R, Lecchi A, Artoni A . Prevalence of disease and relationships between laboratory phenotype and bleeding severity in platelet primary secretion defects. PLoS One. 2013; 8(4):e60396. PMC: 3614926. DOI: 10.1371/journal.pone.0060396. View

3.
Aranda E, Pereira J, Ajenjo C, Prieto C, Sepulveda S, Mezzano D . Human intraplatelet 5-hydroxytryptamine is correlated with mean platelet survival time. Thromb Res. 1996; 84(1):67-72. DOI: 10.1016/0049-3848(96)00162-4. View

4.
Quiroga T, Goycoolea M, Panes O, Aranda E, Martinez C, Belmont S . High prevalence of bleeders of unknown cause among patients with inherited mucocutaneous bleeding. A prospective study of 280 patients and 299 controls. Haematologica. 2007; 92(3):357-65. DOI: 10.3324/haematol.10816. View

5.
Mombouli J, Vanhoutte P . Endothelial dysfunction: from physiology to therapy. J Mol Cell Cardiol. 1999; 31(1):61-74. DOI: 10.1006/jmcc.1998.0844. View