Value of Deep Learning Reconstruction of Chest Low-dose CT for Image Quality Improvement and Lung Parenchyma Assessment on Lung Window

Overview
Journal Eur Radiol
Specialty Radiology
Date 2023 Aug 15
PMID 37581663
Authors
Affiliations
Soon will be listed here.
Abstract

Objectives: To explore the performance of low-dose computed tomography (LDCT) with deep learning reconstruction (DLR) for the improvement of image quality and assessment of lung parenchyma.

Methods: Sixty patients underwent chest regular-dose CT (RDCT) followed by LDCT during the same examination. RDCT images were reconstructed with hybrid iterative reconstruction (HIR) and LDCT images were reconstructed with HIR and DLR, both using lung algorithm. Radiation exposure was recorded. Image noise, signal-to-noise ratio, and subjective image quality of normal and abnormal CT features were evaluated and compared using the Kruskal-Wallis test with Bonferroni correction.

Results: The effective radiation dose of LDCT was significantly lower than that of RDCT (0.29 ± 0.03 vs 2.05 ± 0.65 mSv, p < 0.001). The mean image noise ± standard deviation was 33.9 ± 4.7, 39.6 ± 4.3, and 31.1 ± 3.2 HU in RDCT, LDCT HIR-Strong, and LDCT DLR-Strong, respectively (p < 0.001). The overall image quality of LDCT DLR-Strong was significantly better than that of LDCT HIR-Strong (p < 0.001) and comparable to that of RDCT (p > 0.05). LDCT DLR-Strong was comparable to RDCT in evaluating solid nodules, increased attenuation, linear opacity, and airway lesions (all p > 0.05). The visualization of subsolid nodules and decreased attenuation was better with DLR than with HIR in LDCT but inferior to RDCT (all p < 0.05).

Conclusion: LDCT DLR can effectively reduce image noise and improve image quality. LDCT DLR provides good performance for evaluating pulmonary lesions, except for subsolid nodules and decreased lung attenuation, compared to RDCT-HIR.

Clinical Relevance Statement: The study prospectively evaluated the contribution of DLR applied to chest low-dose CT for image quality improvement and lung parenchyma assessment. DLR can be used to reduce radiation dose and keep image quality for several indications.

Key Points: • DLR enables LDCT maintaining image quality even with very low radiation doses. • Chest LDCT with DLR can be used to evaluate lung parenchymal lesions except for subsolid nodules and decreased lung attenuation. • Diagnosis of pulmonary emphysema or subsolid nodules may require higher radiation doses.

Citing Articles

Ultra-low dose dual-layer detector spectral CT for pulmonary nodule screening: image quality and diagnostic performance.

Ding L, Chen M, Li X, Wu Y, Li J, Deng S Insights Imaging. 2025; 16(1):11.

PMID: 39792229 PMC: 11723867. DOI: 10.1186/s13244-024-01888-1.


Deep learning model for automated diagnosis of moyamoya disease based on magnetic resonance angiography.

Lu M, Zheng Y, Liu S, Zhang X, Lv J, Liu Y EClinicalMedicine. 2024; 77:102888.

PMID: 39559186 PMC: 11570825. DOI: 10.1016/j.eclinm.2024.102888.


Deep Learning-Based Reconstruction Algorithm With Lung Enhancement Filter for Chest CT: Effect on Image Quality and Ground Glass Nodule Sharpness.

Hwang M, Kang S, Lee J, Lee G Korean J Radiol. 2024; 25(9):833-842.

PMID: 39197828 PMC: 11361802. DOI: 10.3348/kjr.2024.0472.


Influence of deep learning image reconstruction algorithm for reducing radiation dose and image noise compared to iterative reconstruction and filtered back projection for head and chest computed tomography examinations: a systematic review.

Chandran M O, Pendem S, P S P, Chacko C, - P, Kadavigere R F1000Res. 2024; 13:274.

PMID: 38725640 PMC: 11079581. DOI: 10.12688/f1000research.147345.1.

References
1.
Smith-Bindman R, Wang Y, Chu P, Chung R, Einstein A, Balcombe J . International variation in radiation dose for computed tomography examinations: prospective cohort study. BMJ. 2019; 364:k4931. PMC: 6314083. DOI: 10.1136/bmj.k4931. View

2.
Kim H, Park C, Chae H, Lee S, Goo J . Impact of radiation dose and iterative reconstruction on pulmonary nodule measurements at chest CT: a phantom study. Diagn Interv Radiol. 2015; 21(6):459-65. PMC: 4622392. DOI: 10.5152/dir.2015.14541. View

3.
Zhang M, Qi W, Sun Y, Jiang Y, Liu X, Hong N . Screening for lung cancer using sub-millisievert chest CT with iterative reconstruction algorithm: image quality and nodule detectability. Br J Radiol. 2017; 91(1090):20170658. PMC: 6350471. DOI: 10.1259/bjr.20170658. View

4.
Ye K, Zhu Q, Li M, Lu Y, Yuan H . A feasibility study of pulmonary nodule detection by ultralow-dose CT with adaptive statistical iterative reconstruction-V technique. Eur J Radiol. 2019; 119:108652. DOI: 10.1016/j.ejrad.2019.108652. View

5.
Willemink M, Noel P . The evolution of image reconstruction for CT-from filtered back projection to artificial intelligence. Eur Radiol. 2018; 29(5):2185-2195. PMC: 6443602. DOI: 10.1007/s00330-018-5810-7. View