Comparison of Reverse Torque Values of Abutment Screws with the Application of Oil-based and Water-based Antibacterial Agents

Overview
Specialty Dentistry
Date 2023 Aug 10
PMID 37560492
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Using antibacterial agents to remove the foul odor of the implant cavity and prevent peri-implantitis can affect the detorque values and lead to the loosening of the abutment screw. This study investigated the effects of tetracycline and chlorhexidine gel on detorque values.

Methods: This in vitro study was carried out on three groups of five implants. Group G1 was the control group, and no material was applied to the implant cavity. In group G2, the implant cavity was first filled with artificial saliva and then with chlorhexidine gel. In group G3, the implant cavity was first filled with artificial saliva and then with tetracycline. The abutments were tightened with 25 N/cm and then loosened. Finally, the detorque values were calculated.

Results: The highest detorque values were recorded in group G1. Group G3 showed the lowest detorque values. ANOVA showed significant differences in mean detorque values (<0.05) between the three groups.

Conclusion: According to this study, applying antibacterial agents decreased the detorque values and increased the risk of screw loosening. The reduction of detorque values was more pronounced with the oil-based antibacterial agent (tetracycline).

Citing Articles

Effects of Blood Contamination and Decontamination Protocol on Reverse Torque Value of Abutment Screws in Dental Implants: An In Vitro Study.

Adawi H, Dewan H, Khawaji A, Akkam H, Hakami A, Wasli B Biomimetics (Basel). 2023; 8(2).

PMID: 37092409 PMC: 10123743. DOI: 10.3390/biomimetics8020157.

References
1.
Tan K, Nicholls J . The effect of 3 torque delivery systems on gold screw preload at the gold cylinder-abutment screw joint. Int J Oral Maxillofac Implants. 2002; 17(2):175-83. View

2.
Theoharidou A, Petridis H, Tzannas K, Garefis P . Abutment screw loosening in single-implant restorations: a systematic review. Int J Oral Maxillofac Implants. 2008; 23(4):681-90. View

3.
Broggini N, McManus L, Hermann J, Medina R, Schenk R, Buser D . Peri-implant inflammation defined by the implant-abutment interface. J Dent Res. 2006; 85(5):473-8. DOI: 10.1177/154405910608500515. View

4.
Paepoemsin T, Reichart P, Chaijareenont P, Strietzel F, Khongkhunthian P . Removal torque evaluation of three different abutment screws for single implant restorations after mechanical cyclic loading. Oral Implantol (Rome). 2017; 9(4):213-221. PMC: 5159936. DOI: 10.11138/orl/2016.9.4.213. View

5.
Grunder U, Gracis S, Capelli M . Influence of the 3-D bone-to-implant relationship on esthetics. Int J Periodontics Restorative Dent. 2005; 25(2):113-9. View