Transcatheter aortic valve replacement (TAVR) is a treatment option for patients with symptomatic severe aortic stenosis across the entire spectrum of surgical risk. Recent trial data have led to the expansion of TAVR into lower-risk patients. With iterative technological advances and successive increases in procedural experience, the occurrence of complications following TAVR has declined. One of the most feared complications remains stroke, and patients consider stroke a worse outcome than death. There has therefore been great interest in strategies to mitigate the risk of stroke in patients undergoing TAVR. In this paper, we will discuss mechanisms and predictors of stroke after TAVR and describe the currently available cerebral embolic protection devices, including their design and relevant clinical studies pertaining to their use. We will also review the current overall evidence base for cerebral embolic protection during TAVR and ongoing randomized controlled trials. Finally, we will discuss our pragmatic recommendations for the use of cerebral embolic protection devices in patients undergoing TAVR.

Citing Articles

Performance Assessment of an Electrostatic Filter-Diverter Stent Cerebrovascular Protection Device: Evaluation of a Range of Potential Electrostatic Fields Focusing on Small Particles.

Eguzkitza B, Navia J, Houzeaux G, Butakoff C, Vazquez M Bioengineering (Basel). 2024; 11(11).

PMID: 39593787 PMC: 11592227. DOI: 10.3390/bioengineering11111127.


Association of Self-Expanding Valves with Increased Capture of Larger Debris by Cerebral Protection System in Transcatheter Aortic Valve Replacement for Severe Aortic Stenosis.

Ling A, Lee Y, Yin W, Tsao T, Lee K, Lin H Acta Cardiol Sin. 2024; 40(6):751-761.

PMID: 39582846 PMC: 11579694. DOI: 10.6515/ACS.202411_40(6).20240730A.


Cerebral Embolic Protection Devices: Are There Any Indications in Transcatheter Aortic Valve Replacement?.

Sayah N, Skalidis I, Mesnier J, Neylon A, Akodad M, Asgar A J Clin Med. 2024; 13(18).

PMID: 39336957 PMC: 11432207. DOI: 10.3390/jcm13185471.


A Simple Technique for Deploying the SENTINEL Cerebral Protection System in Bovine Aortic Arch Anatomy.

Kattoor A, Manion C, Iyer V Struct Heart. 2024; 8(2):100228.

PMID: 38481710 PMC: 10927443. DOI: 10.1016/j.shj.2023.100228.

References
1.
Makkar R, Fontana G, Sondergaard L . Possible Subclinical Leaflet Thrombosis in Bioprosthetic Aortic Valves. N Engl J Med. 2016; 374(16):1591-2. DOI: 10.1056/NEJMc1600179. View

2.
Kapadia S, Kodali S, Makkar R, Mehran R, Lazar R, Zivadinov R . Protection Against Cerebral Embolism During Transcatheter Aortic Valve Replacement. J Am Coll Cardiol. 2016; 69(4):367-377. DOI: 10.1016/j.jacc.2016.10.023. View

3.
Nazif T, Moses J, Sharma R, Dhoble A, Rovin J, Brown D . Randomized Evaluation of TriGuard 3 Cerebral Embolic Protection After Transcatheter Aortic Valve Replacement: REFLECT II. JACC Cardiovasc Interv. 2021; 14(5):515-527. DOI: 10.1016/j.jcin.2020.11.011. View

4.
Megaly M, Sorajja P, Cavalcante J, Pershad A, Gossl M, Abraham B . Ischemic Stroke With Cerebral Protection System During Transcatheter Aortic Valve Replacement. JACC Cardiovasc Interv. 2020; 13(18):2149-2155. DOI: 10.1016/j.jcin.2020.05.039. View

5.
Van Mieghem N, Schipper M, Ladich E, Faqiri E, van der Boon R, Randjgari A . Histopathology of embolic debris captured during transcatheter aortic valve replacement. Circulation. 2013; 127(22):2194-201. DOI: 10.1161/CIRCULATIONAHA.112.001091. View