Novel Multi-Parametric Sensor System for Comprehensive Multi-Wavelength Photoplethysmography Characterization

Overview
Journal Sensors (Basel)
Publisher MDPI
Specialty Biotechnology
Date 2023 Jul 29
PMID 37514922
Authors
Affiliations
Soon will be listed here.
Abstract

Photoplethysmography (PPG) is widely used to assess cardiovascular health. However, its usage and standardization are limited by the impact of variable contact force and temperature, which influence the accuracy and reliability of the measurements. Although some studies have evaluated the impact of these phenomena on signal amplitude, there is still a lack of knowledge about how these perturbations can distort the signal morphology, especially for multi-wavelength PPG (MW-PPG) measurements. This work presents a modular multi-parametric sensor system that integrates continuous and real-time acquisition of MW-PPG, contact force, and temperature signals. The implemented design solution allows for a comprehensive characterization of the effects of the variations in these phenomena on the contour of the MW-PPG signal. Furthermore, a dynamic DC cancellation circuitry was implemented to improve measurement resolution and obtain high-quality raw multi-parametric data. The accuracy of the MW-PPG signal acquisition was assessed using a synthesized reference PPG optical signal. The performance of the contact force and temperature sensors was evaluated as well. To determine the overall quality of the multi-parametric measurement, an in vivo measurement on the index finger of a volunteer was performed. The results indicate a high precision and accuracy in the measurements, wherein the capacity of the system to obtain high-resolution and low-distortion MW-PPG signals is highlighted. These findings will contribute to developing new signal-processing approaches, advancing the accuracy and robustness of PPG-based systems, and bridging existing gaps in the literature.

Citing Articles

Exploring Near- and Far-Field Effects in Photoplethysmography Signals Across Different Source-Detector Distances.

Sole Morillo A, Cause J, De Pauw K, Silva B, Stiens J Sensors (Basel). 2025; 25(1.

PMID: 39796889 PMC: 11722670. DOI: 10.3390/s25010099.


Evaluating Vascular Depth-Dependent Changes in Multi-Wavelength PPG Signals Due to Contact Force.

Cause J, Sole Morillo A, Silva B, Garcia-Naranjo J, Stiens J Sensors (Basel). 2024; 24(9).

PMID: 38732798 PMC: 11085639. DOI: 10.3390/s24092692.


Anomaly Detection in Multi-Wavelength Photoplethysmography Using Lightweight Machine Learning Algorithms.

Baciu V, Cause J, Sole Morillo A, Garcia-Naranjo J, Stiens J, Silva B Sensors (Basel). 2023; 23(15).

PMID: 37571730 PMC: 10422657. DOI: 10.3390/s23156947.

References
1.
Chandrasekhar A, Yavarimanesh M, Natarajan K, Hahn J, Mukkamala R . PPG Sensor Contact Pressure Should Be Taken Into Account for Cuff-Less Blood Pressure Measurement. IEEE Trans Biomed Eng. 2020; 67(11):3134-3140. PMC: 8856571. DOI: 10.1109/TBME.2020.2976989. View

2.
Teng X, Zhang Y . Theoretical study on the effect of sensor contact force on pulse transit time. IEEE Trans Biomed Eng. 2007; 54(8):1490-8. DOI: 10.1109/TBME.2007.900815. View

3.
Moraes J, Rocha M, Vasconcelos G, Vasconcelos Filho J, de Albuquerque V, Alexandria A . Advances in Photopletysmography Signal Analysis for Biomedical Applications. Sensors (Basel). 2018; 18(6). PMC: 6022166. DOI: 10.3390/s18061894. View

4.
Roach J, Thiele R . Perioperative blood pressure monitoring. Best Pract Res Clin Anaesthesiol. 2019; 33(2):127-138. DOI: 10.1016/j.bpa.2019.05.001. View

5.
Lee J, Yang S, Park J, Kim H, Kim E, Jang Y . Time to consider the contact force during photoplethysmography measurement during pediatric anesthesia: A prospective, nonrandomized interventional study. Paediatr Anaesth. 2018; 28(7):660-667. DOI: 10.1111/pan.13415. View