» Articles » PMID: 37419940

Application of Error Classification Model Using Indices Based on Dose Distribution for Characteristics Evaluation of Multileaf Collimator Position Errors

Overview
Journal Sci Rep
Specialty Science
Date 2023 Jul 7
PMID 37419940
Authors
Affiliations
Soon will be listed here.
Abstract

This study aims to evaluate the specific characteristics of various multileaf collimator (MLC) position errors that are correlated with the indices using dose distribution. The dose distribution was investigated using the gamma, structural similarity, and dosiomics indices. Cases from the American Association of Physicists in Medicine Task Group 119 were planned, and systematic and random MLC position errors were simulated. The indices were obtained from distribution maps and statistically significant indices were selected. The final model was determined when all values of the area under the curve, accuracy, precision, sensitivity, and specificity were higher than 0.8 (p < 0.05). The dose-volume histogram (DVH) relative percentage difference between the error-free and error datasets was examined to investigate clinical relations. Seven multivariate predictive models were finalized. The common significant dosiomics indices (GLCM Energy and GLRLM_LRHGE) can characterize the MLC position error. In addition, the finalized logistic regression model for MLC position error prediction showed excellent performance with AUC > 0.9. Furthermore, the results of the DVH were related to dosiomics analysis in that it reflects the characteristics of the MLC position error. It was also shown that dosiomics analysis could provide important information on localized dose-distribution differences in addition to DVH information.

References
1.
Ezzell G, Burmeister J, Dogan N, LoSasso T, Mechalakos J, Mihailidis D . IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med Phys. 2009; 36(11):5359-73. DOI: 10.1118/1.3238104. View

2.
Moradmand H, Aghamiri S, Ghaderi R . Impact of image preprocessing methods on reproducibility of radiomic features in multimodal magnetic resonance imaging in glioblastoma. J Appl Clin Med Phys. 2019; 21(1):179-190. PMC: 6964771. DOI: 10.1002/acm2.12795. View

3.
Zhen H, Nelms B, Tome W . Moving from gamma passing rates to patient DVH-based QA metrics in pretreatment dose QA. Med Phys. 2011; 38(10):5477-89. DOI: 10.1118/1.3633904. View

4.
Kadoya N, Kon Y, Takayama Y, Matsumoto T, Hayashi N, Katsuta Y . Quantifying the performance of two different types of commercial software programs for 3D patient dose reconstruction for prostate cancer patients: Machine log files vs. machine log files with EPID images. Phys Med. 2018; 45:170-176. DOI: 10.1016/j.ejmp.2017.12.018. View

5.
Stevens M, Lobb E, Yenice K . Validation of MLC-based linac radiosurgery for trigeminal neuralgia. J Appl Clin Med Phys. 2018; 19(4):214-221. PMC: 6036389. DOI: 10.1002/acm2.12381. View