Quantitative Dual-energy Computed Tomography with Cesium As a Novel Contrast Agent for Localization of Thermochemical Ablation in Phantoms and Ex Vivo Models

Overview
Journal Med Phys
Specialty Biophysics
Date 2023 Jul 6
PMID 37409792
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Thermochemical ablation (TCA) is a minimally invasive therapy under development for hepatocellular carcinoma. TCA simultaneously delivers an acid (acetic acid, AcOH) and base (sodium hydroxide, NaOH) directly into the tumor, where the acid/base chemical reaction produces an exotherm that induces local ablation. However, AcOH and NaOH are not radiopaque, making monitoring TCA delivery difficult.

Purpose: We address the issue of image guidance for TCA by utilizing cesium hydroxide (CsOH) as a novel theranostic component of TCA that is detectable and quantifiable with dual-energy CT (DECT).

Materials And Methods: To quantify the minimum concentration of CsOH that can be positively identified by DECT, the limit of detection (LOD) was established in an elliptical phantom (Multi-Energy CT Quality Assurance Phantom, Kyoto Kagaku, Kyoto, Japan) with two DECT technologies: a dual-source system (SOMATOM Force, Siemens Healthineers, Forchheim, Germany) and a split-filter, single-source system (SOMATOM Edge, Siemens Healthineers). The dual-energy ratio (DER) and LOD of CsOH were determined for each system. Cesium concentration quantification accuracy was evaluated in a gelatin phantom before quantitative mapping was performed in ex vivo models.

Results: On the dual-source system, the DER and LOD were 2.94 and 1.36-mM CsOH, respectively. For the split-filter system, the DER and LOD were 1.41- and 6.11-mM CsOH, respectively. The signal on cesium maps in phantoms tracked linearly with concentration (R  = 0.99) on both systems with an RMSE of 2.56 and 6.72 on the dual-source and split-filter system, respectively. In ex vivo models, CsOH was detected following delivery of TCA at all concentrations.

Conclusions: DECT can be used to detect and quantify the concentration of cesium in phantom and ex vivo tissue models. When incorporated in TCA, CsOH performs as a theranostic agent for quantitative DECT image-guidance.

References
1.
Sessions D, Heard K, Kosnett M . Fatal cesium chloride toxicity after alternative cancer treatment. J Altern Complement Med. 2013; 19(12):973-5. PMC: 3868249. DOI: 10.1089/acm.2012.0745. View

2.
Dimitroulis D, Damaskos C, Valsami S, Davakis S, Garmpis N, Spartalis E . From diagnosis to treatment of hepatocellular carcinoma: An epidemic problem for both developed and developing world. World J Gastroenterol. 2017; 23(29):5282-5294. PMC: 5550777. DOI: 10.3748/wjg.v23.i29.5282. View

3.
Sheth R, Hesketh R, Kong D, Wicky S, Oklu R . Barriers to drug delivery in interventional oncology. J Vasc Interv Radiol. 2013; 24(8):1201-7. DOI: 10.1016/j.jvir.2013.03.034. View

4.
Kelcz F, Joseph P, Hilal S . Noise considerations in dual energy CT scanning. Med Phys. 1979; 6(5):418-25. DOI: 10.1118/1.594520. View

5.
Perez J, Jacobsen M, Damasco J, Melancon A, Huang S, Layman R . Optimization of the differentiation and quantification of high-Z nanoparticles incorporated in medical devices for CT-guided interventions. Med Phys. 2020; 48(1):300-312. PMC: 8097741. DOI: 10.1002/mp.14601. View