Enhancing Radiation Dose Efficiency in Prospective ECG-Triggered Coronary CT Angiography Using Calcium-Scoring CT

Overview
Specialty Radiology
Date 2023 Jun 28
PMID 37370957
Authors
Affiliations
Soon will be listed here.
Abstract

Background: This study investigates whether the scan length adjustment of prospectively ECG-triggered coronary CT angiography (CCTA) using calcium-scoring CT (CAS-CT) images can reduce overall radiation doses.

Methods: A retrospective analysis was conducted on 182 patients who underwent CAS-CT and prospectively ECG-triggered CCTA using a second-generation Dual-Source CT scanner. CCTA planning was based on CAS-CT images, for which simulated scout view planning was performed for comparison. Effective doses were compared between two scenarios: Scenario 1-CAS-CT-derived CCTA + CAS-CT and Scenario 2-scout-view-derived CCTA without CAS-CT. Dose differences were further analyzed with respect to scan mode and body mass index.

Results: Planning CCTA using CAS-CT led to a shorter scan length than planning via scout view (114.3 ± 9.7 mm vs. 133.7 ± 13.2 mm, < 0.001). The whole-examination effective dose was slightly lower for Scenario 1 (3.2 [1.8-5.3] mSv vs. 3.4 [1.5-5.9] mSv; < 0.001, = 182). Notably, Scenario 1 resulted in a significantly lower radiation dose for sequential scans and obese patients. Only high-pitch spiral CCTA showed dose reduction in Scenario 2.

Conclusions: Using CAS-CT for planning prospectively ECG-triggered CCTA reduced the overall radiation dose administered compared to scout view planning without CAS-CT, except for high-pitch spiral CCTA, where a slightly opposite effect was observed.

Citing Articles

Achieving low radiation dose and contrast agents dose in coronary CT angiography at 60-kVp ultra-low tube voltage.

He W, Huang F, Wu X, Xie A, Sun W, Liu P BMC Med Imaging. 2025; 25(1):73.

PMID: 40038586 PMC: 11877812. DOI: 10.1186/s12880-025-01608-1.


Evaluating small coronary stents with dual-source photon-counting computed tomography: effect of different scan modes on image quality and performance in a phantom.

Stein T, von Zur Muhlen C, Verloh N, Schurmann T, Krauss T, Soschynski M Diagn Interv Radiol. 2024; 31(1):29-38.

PMID: 39463047 PMC: 11701693. DOI: 10.4274/dir.2024.242893.

References
1.
Leschka S, Alkadhi H, Plass A, Desbiolles L, Grunenfelder J, Marincek B . Accuracy of MSCT coronary angiography with 64-slice technology: first experience. Eur Heart J. 2005; 26(15):1482-7. DOI: 10.1093/eurheartj/ehi261. View

2.
Abada H, Larchez C, Daoud B, Sigal-Cinqualbre A, Paul J . MDCT of the coronary arteries: feasibility of low-dose CT with ECG-pulsed tube current modulation to reduce radiation dose. AJR Am J Roentgenol. 2006; 186(6 Suppl 2):S387-90. DOI: 10.2214/AJR.05.0216. View

3.
Cury R, Leipsic J, Abbara S, Achenbach S, Berman D, Bittencourt M . CAD-RADS™ 2.0 - 2022 Coronary Artery Disease-Reporting and Data System: An Expert Consensus Document of the Society of Cardiovascular Computed Tomography (SCCT), the American College of Cardiology (ACC), the American College of Radiology (ACR), and.... J Cardiovasc Comput Tomogr. 2022; 16(6):536-557. DOI: 10.1016/j.jcct.2022.07.002. View

4.
Soschynski M, Hagar M, Taron J, Krauss T, Ruile P, Hein M . Update for the Performance of CT Coronary Angiography - Evidence-Based Application and Technical Guidance According to Current Consensus Guidelines and Practical Advice from the Clinical Routine. Rofo. 2022; 194(6):613-624. DOI: 10.1055/a-1747-3554. View

5.
Halliburton S, Abbara S, Chen M, Gentry R, Mahesh M, Raff G . SCCT guidelines on radiation dose and dose-optimization strategies in cardiovascular CT. J Cardiovasc Comput Tomogr. 2011; 5(4):198-224. PMC: 3391026. DOI: 10.1016/j.jcct.2011.06.001. View