Simulation of a New Respiratory Phase Sorting Method for 4D-imaging Using Optical Surface Information Towards Precision Radiotherapy

Overview
Journal Comput Biol Med
Publisher Elsevier
Date 2023 Jun 8
PMID 37290392
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Respiratory signal detection is critical for 4-dimensional (4D) imaging. This study proposes and evaluates a novel phase sorting method using optical surface imaging (OSI), aiming to improve the precision of radiotherapy.

Method: Based on 4D Extended Cardiac-Torso (XCAT) digital phantom, OSI in point cloud format was generated from the body segmentation, and image projections were simulated using the geometries of Varian 4D kV cone-beam-CT (CBCT). Respiratory signals were extracted respectively from the segmented diaphragm image (reference method) and OSI respectively, where Gaussian Mixture Model and Principal Component Analysis (PCA) were used for image registration and dimension reduction respectively. Breathing frequencies were compared using Fast-Fourier-Transform. Consistency of 4DCBCT images reconstructed using Maximum Likelihood Expectation Maximization algorithm was also evaluated quantitatively, where high consistency can be suggested by lower Root-Mean-Square-Error (RMSE), Structural-Similarity-Index (SSIM) value closer to 1, and larger Peak-Signal-To-Noise-Ratio (PSNR) respectively.

Results: High consistency of breathing frequencies was observed between the diaphragm-based (0.232 Hz) and OSI-based (0.251 Hz) signals, with a slight discrepancy of 0.019Hz. Using end of expiration (EOE) and end of inspiration (EOI) phases as examples, the mean±1SD values of the 80 transverse, 100 coronal and 120 sagittal planes were 0.967, 0,972, 0.974 (SSIM); 1.657 ± 0.368, 1.464 ± 0.104, 1.479 ± 0.297 (RMSE); and 40.501 ± 1.737, 41.532 ± 1.464, 41.553 ± 1.910 (PSNR) for the EOE; and 0.969, 0.973, 0.973 (SSIM); 1.686 ± 0.278, 1.422 ± 0.089, 1.489 ± 0.238 (RMSE); and 40.535 ± 1.539, 41.605 ± 0.534, 41.401 ± 1.496 (PSNR) for EOI respectively.

Conclusions: This work proposed and evaluated a novel respiratory phase sorting approach for 4D imaging using optical surface signals, which can potentially be applied to precision radiotherapy. Its potential advantages were non-ionizing, non-invasive, non-contact, and more compatible with various anatomic regions and treatment/imaging systems.

Citing Articles

Liver respiratory-induced motion estimation using abdominal surface displacement as a surrogate: robotic phantom and clinical validation with varied correspondence models.

Cordon Avila A, Abayazid M Int J Comput Assist Radiol Surg. 2024; 19(8):1477-1487.

PMID: 38809319 PMC: 11329552. DOI: 10.1007/s11548-024-03176-1.


Surrogate-driven respiratory motion model for projection-resolved motion estimation and motion compensated cone-beam CT reconstruction from unsorted projection data.

Huang Y, Thielemans K, Price G, McClelland J Phys Med Biol. 2023; 69(2).

PMID: 38091611 PMC: 10791594. DOI: 10.1088/1361-6560/ad1546.

References
1.
Chu M, Nguyen T, Pandey V, Zhou Y, Pham H, Bar-Yoseph R . Respiration rate and volume measurements using wearable strain sensors. NPJ Digit Med. 2019; 2:8. PMC: 6550208. DOI: 10.1038/s41746-019-0083-3. View

2.
Ehman R, McNamara M, PALLACK M, Hricak H, Higgins C . Magnetic resonance imaging with respiratory gating: techniques and advantages. AJR Am J Roentgenol. 1984; 143(6):1175-82. DOI: 10.2214/ajr.143.6.1175. View

3.
Vedam S, Kini V, Keall P, Ramakrishnan V, Mostafavi H, Mohan R . Quantifying the predictability of diaphragm motion during respiration with a noninvasive external marker. Med Phys. 2003; 30(4):505-13. DOI: 10.1118/1.1558675. View

4.
Zhou L, Schaefferkoetter J, Tham I, Huang G, Yan J . Supervised learning with cyclegan for low-dose FDG PET image denoising. Med Image Anal. 2020; 65:101770. DOI: 10.1016/j.media.2020.101770. View

5.
Siegel R, Miller K, Jemal A . Cancer statistics, 2020. CA Cancer J Clin. 2020; 70(1):7-30. DOI: 10.3322/caac.21590. View