» Articles » PMID: 3698394

The Problem of Surface Damage in Polyethylene Total Knee Components

Overview
Publisher Wolters Kluwer
Specialty Orthopedics
Date 1986 Apr 1
PMID 3698394
Citations 13
Authors
Affiliations
Soon will be listed here.
Abstract

Observations of surface damage on retrieved total knee polyethylene components have been combined with experimental and analytical studies of the contact problem to identify important clinical and design factors. The amount and severity of damage occurring on the articulating surfaces of knee components increases significantly with patient weight and with the length of time the component is implanted. The design variables that affect the amount of damage are component thickness, the conformity of the articulating surfaces, and the type of polyethylene material used. Surface damage will be more severe in thin (less than 4-6 mm) components and in components with relatively flat tibial articulating surfaces. Surface damage is also expected to be more extreme for carbon-reinforced polyethylene components than for components made from plain polyethylene.

Citing Articles

Gait analysis methodology for the measurement of biomechanical parameters in total knee arthroplasties. A literature review.

Papagiannis G, Triantafyllou A, Roumpelakis I, Papagelopoulos P, Babis G J Orthop. 2018; 15(1):181-185.

PMID: 29657464 PMC: 5895909. DOI: 10.1016/j.jor.2018.01.048.


Polyethylene Damage Increases With Varus Implant Alignment in Posterior-stabilized and Constrained Condylar Knee Arthroplasty.

Li Z, Esposito C, Koch C, Lee Y, Padgett D, Wright T Clin Orthop Relat Res. 2017; 475(12):2981-2991.

PMID: 28822068 PMC: 5670063. DOI: 10.1007/s11999-017-5477-4.


Does impact sport activity influence total hip arthroplasty durability?.

Ollivier M, Frey S, Parratte S, Flecher X, Argenson J Clin Orthop Relat Res. 2012; 470(11):3060-6.

PMID: 22535588 PMC: 3462849. DOI: 10.1007/s11999-012-2362-z.


History and systematic review of wear and osteolysis outcomes for first-generation highly crosslinked polyethylene.

Kurtz S, Gawel H, Patel J Clin Orthop Relat Res. 2011; 469(8):2262-77.

PMID: 21431461 PMC: 3126942. DOI: 10.1007/s11999-011-1872-4.


What have we learned from LCS mobile-bearing knee system?.

Cho W, Youm Y, Ahn S, Sohn D Knee Surg Sports Traumatol Arthrosc. 2010; 18(10):1345-51.

PMID: 20556357 DOI: 10.1007/s00167-010-1166-5.