Inter-observer and Intra-observer Reliability Between Manual Segmentation and Semi-automated Segmentation for Carotid Vessel Wall Volume Measurements on Three-dimensional Ultrasonography

Overview
Journal Ultrasonography
Date 2023 Mar 20
PMID 36935603
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Carotid vessel wall volume (VWV) measurement on three-dimensional ultrasonography (3DUS) outperforms conventional two-dimensional ultrasonography for carotid atherosclerosis evaluation. Although time-saving semi-automated algorithms have been introduced, their clinical availability remains limited due to a lack of validation, particularly an extensive reliability analysis. This study compared inter-observer and intra-observer reliability between manual segmentation and semi-automated segmentation for carotid VWV measurements on 3DUS.

Methods: Thirty-one 3DUS volume datasets were prospectively acquired from 20 healthy subjects, aged >18 years, without previous stroke, transient ischemic attack, or cardiovascular disease. Five observers segmented all volume datasets both manually and semi-automatically. The process was repeated five times. Reliability was expressed by the intraclass correlation coefficient, supplemented by the coefficient of variation.

Results: Carotid VWV measurements using the common carotid artery (CCA) were more reliable than those using the internal carotid artery (ICA) or external carotid artery (ECA) for both manual and semiautomated segmentation (manual segmentation, CCA: inter-observer, 0.935; intra-observer, 0.934 to 0.966; ICA: inter-observer, 0.784; intra-observer, 0.756 to 0.878; ECA: inter-observer, 0.732; intraobserver, 0.919 to 0.962; semi-automated segmentation, CCA: inter-observer, 0.986; intra-observer, 0.954 to 0.993; ICA: inter-observer, 0.977; intra-observer, 0.958 to 0.978; ECA: inter-observer, 0.966; intra-observer, 0.884 to 0.937). Total carotid VWV measurements by manual (inter-observer, 0.922; intra-observer, 0.927 to 0.961) and semi-automated segmentation (inter-observer, 0.987; intra-observer, 0.968 to 0.989) were highly reliable. Semi-automated segmentation showed higher reliability than manual segmentation for both individual and total carotid VWV measurements.

Conclusion: 3DUS carotid VWV measurements of the CCA are more reliable than measurements of the ICA and ECA. Total carotid VWV measurements are highly reliable. Semi-automated segmentation has higher reliability than manual segmentation.

Citing Articles

Interobserver Variability in Manual Versus Semi-Automatic CT Assessments of Small Lung Nodule Diameter and Volume.

Zacharias F, Svahn T Tomography. 2024; 10(12):2087-2099.

PMID: 39728910 PMC: 11680079. DOI: 10.3390/tomography10120148.

References
1.
Browne J, Watson A, Hoskins P, Elliott A . Investigation of the effect of subcutaneous fat on image quality performance of 2D conventional imaging and tissue harmonic imaging. Ultrasound Med Biol. 2005; 31(7):957-64. DOI: 10.1016/j.ultrasmedbio.2005.03.012. View

2.
Kobayashi K, Kishi M, Atsumi T, Bertolaccini M, Makino H, Sakairi N . Circulating oxidized LDL forms complexes with beta2-glycoprotein I: implication as an atherogenic autoantigen. J Lipid Res. 2003; 44(4):716-26. DOI: 10.1194/jlr.M200329-JLR200. View

3.
Roger V, Go A, Lloyd-Jones D, Benjamin E, Berry J, Borden W . Heart disease and stroke statistics--2012 update: a report from the American Heart Association. Circulation. 2011; 125(1):e2-e220. PMC: 4440543. DOI: 10.1161/CIR.0b013e31823ac046. View

4.
Heliopoulos J, Vadikolias K, Piperidou C, Mitsias P . Detection of carotid artery plaque ulceration using 3-dimensional ultrasound. J Neuroimaging. 2009; 21(2):126-31. DOI: 10.1111/j.1552-6569.2009.00450.x. View

5.
Goldstein A . Errors in ultrasound digital image distance measurements. Ultrasound Med Biol. 2000; 26(7):1125-32. DOI: 10.1016/s0301-5629(00)00249-0. View