Bias and Precision of Cholesterol Analysis by Physician's Office Analyzers

Overview
Journal Clin Chem
Specialty Biochemistry
Date 1987 Dec 1
PMID 3690842
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

We studied the bias and precision of serum cholesterol analysis by physician's office analyzers. Total imprecision (CV range, %) for analysis of serum pools was: Abbott Vision 1.5%-1.9%; Ames Seralyzer 3.9%-4.5%; BMD Reflotron 2.3%-3.8%; Chrometrics Cholesterol Test System 2.3%-2.8%; Kodak DT-60 1.6%-2.2%. The Ames Seralyzer exhibited an excessive between-run component of variation. We collected, from 109 volunteers, samples of venous serum, heparin-treated whole blood, heparin-treated plasma, and fingerstick whole blood, and analyzed each type (where possible) with each system; serum was also analyzed in duplicate (by a proposed Reference Method) at the Centers for Disease Control. For assays with serum, the BMD Reflotron and Kodak DT-60 exhibited negative bias. All systems gave lower results for plasma and whole blood than for serum. All systems except the Kodak DT-60 were less precise for analysis of patients' sera than for analysis of serum pools; between-specimen variables may influence the results of these systems.

Citing Articles

Systematic review of near patient test evaluations in primary care.

Delaney B, Hyde C, McManus R, Wilson S, Fitzmaurice D, Jowett S BMJ. 1999; 319(7213):824-7.

PMID: 10496828 PMC: 314212. DOI: 10.1136/bmj.319.7213.824.


Comparison of the use of a dry chemistry analyser in primary care in Norway and the United Kingdom.

Thue G, Sandberg S, Bullock D Br J Gen Pract. 1993; 43(366):10-4.

PMID: 8457356 PMC: 1372200.


Quality of plasma cholesterol measurements in primary care.

Broughton P, Bullock D, Cramb R BMJ. 1989; 298(6669):297-8.

PMID: 2493904 PMC: 1835598. DOI: 10.1136/bmj.298.6669.297.


Precision and accuracy of a portable blood analyzer system during cholesterol screening.

Greenland P, Bowley N, French C, Meiklejohn B, Gagliano S, Sparks C Am J Public Health. 1990; 80(2):181-4.

PMID: 2404422 PMC: 1404611. DOI: 10.2105/ajph.80.2.181.


One state's approach to the regulation of cholesterol screening.

DeBoy J Public Health Rep. 1990; 105(6):584-8.

PMID: 2124359 PMC: 1580175.