» Articles » PMID: 36844719

Patient-specific, Echocardiography Compatible Flow Loop Model of Aortic Valve Regurgitation in the Setting of a Mechanical Assist Device

Overview
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Aortic regurgitation (AR) occurs commonly in patients with continuous-flow left ventricular assist devices (LVAD). No gold standard is available to assess AR severity in this setting. Aim of this study was to create a patient-specific model of AR-LVAD with tailored AR flow assessed by Doppler echocardiography.

Methods: An echo-compatible flow loop incorporating a 3D printed left heart of a Heart Mate II (HMII) recipient with known significant AR was created. Forward flow and LVAD flow at different LVAD speed were directly measured and AR regurgitant volume (RegVol) obtained by subtraction. Doppler parameters of AR were simultaneously measured at each LVAD speed.

Results: We reproduced hemodynamics in a LVAD recipient with AR. AR in the model replicated accurately the AR in the index patient by comparable Color Doppler assessment. Forward flow increased from 4.09 to 5.61 L/min with LVAD speed increasing from 8,800 to 11,000 RPM while RegVol increased by 0.5 L/min (2.01 to 2.5 L/min).

Conclusions: Our circulatory flow loop was able to accurately replicate AR severity and flow hemodynamics in an LVAD recipient. This model can be reliably used to study echo parameters and aid clinical management of patients with LVAD.

Citing Articles

The role of innovative modeling and imaging techniques in improving outcomes in patients with LVAD.

Wilson S, Ingram K, Oh A, Moreno M, Kassi M Front Cardiovasc Med. 2023; 10:1248300.

PMID: 37692033 PMC: 10484111. DOI: 10.3389/fcvm.2023.1248300.

References
1.
Gasparovic H, Kopjar T, Saeed D, cikes M, Svetina L, Petricevic M . De Novo Aortic Regurgitation After Continuous-Flow Left Ventricular Assist Device Implantation. Ann Thorac Surg. 2017; 104(2):704-711. DOI: 10.1016/j.athoracsur.2017.01.114. View

2.
Fried J, Nazif T, Colombo P . A new frontier for TAVR: Aortic insufficiency in CF-LVAD patients. J Heart Lung Transplant. 2019; 38(9):927-929. DOI: 10.1016/j.healun.2019.06.024. View

3.
Grinstein J, Kruse E, Sayer G, Fedson S, Kim G, Jorde U . Accurate Quantification Methods for Aortic Insufficiency Severity in Patients With LVAD: Role of Diastolic Flow Acceleration and Systolic-to-Diastolic Peak Velocity Ratio of Outflow Cannula. JACC Cardiovasc Imaging. 2015; 9(6):641-51. DOI: 10.1016/j.jcmg.2015.06.020. View

4.
Yoganathan A, Tamura T, Tomizuka F, Woo Y, Sahn D . Studies in vitro of the relationship between ultrasound and laser Doppler velocimetry and applicability to the simplified Bernoulli relationship. Circulation. 1986; 73(2):300-8. DOI: 10.1161/01.cir.73.2.300. View

5.
Saeed D, Ootaki Y, Noecker A, Weber S, Smith W, Duncan B . The Cleveland Clinic PediPump: virtual fitting studies in children using three-dimensional reconstructions of cardiac computed tomography scans. ASAIO J. 2008; 54(1):133-7. DOI: 10.1097/MAT.0b013e31815b4495. View