» Articles » PMID: 36811084

A Fully-enclosed Prototype 'pen' for Rapid Detection of SARS-CoV-2 Based on RT-RPA with Dipstick Assay at Point-of-care Testing

Overview
Date 2023 Feb 22
PMID 36811084
Authors
Affiliations
Soon will be listed here.
Abstract

A fully-enclosed prototype 'pen' for rapid detection of SARS-CoV-2 based on reverse transcriptase isothermal recombinase polymerase amplification (RT-RPA) with dipstick assay was developed. The integrated handheld device, consisting of amplification, detection and sealing modules, was developed to perform rapid nucleic acid amplification and detection under a fully enclosed condition. After RT-RPA amplification with a metal bath or a normal PCR instrument, the amplicons were mixed with dilution buffer prior to being detected on a lateral flow strip. To avoid aerosol contamination causing false-positive, from amplification to final detection, the detection 'pen' had been enclosed to isolate from the environment. With colloidal gold strip-based detection, the detection results could be directly observed by eyes. By cooperating with other inexpensive and rapid methods for POC nucleic acid extraction, the developed 'pen' could detect COVID-19 or other infectious diseases in a convenient, simple and reliable way.

Citing Articles

Accurate molecular identification of different meat adulterations without carryover contaminations on a microarray chip PCR-directed microfluidic lateral flow strip device.

Wang H, Meng X, Yao L, Wu Q, Yao B, Chen Z Food Chem (Oxf). 2023; 7:100180.

PMID: 37664158 PMC: 10471925. DOI: 10.1016/j.fochms.2023.100180.

References
1.
Liu W, Zhang M, Liu X, Sharma A, Ding X . A Point-of-Need infrared mediated PCR platform with compatible lateral flow strip for HPV detection. Biosens Bioelectron. 2017; 96:213-219. DOI: 10.1016/j.bios.2017.04.047. View

2.
Wu X, Wang Y, Yin Q, Jiao W, Sun L, Qi H . A diagnostic test that uses isothermal amplification and lateral flow detection sdaA can detect tuberculosis in 60 min. J Appl Microbiol. 2020; 130(6):2102-2110. DOI: 10.1111/jam.14902. View

3.
Mullis K, Faloona F, Scharf S, Saiki R, Horn G, Erlich H . Specific enzymatic amplification of DNA in vitro: the polymerase chain reaction. Cold Spring Harb Symp Quant Biol. 1986; 51 Pt 1:263-73. DOI: 10.1101/sqb.1986.051.01.032. View

4.
Qian J, Boswell S, Chidley C, Lu Z, Pettit M, Gaudio B . An enhanced isothermal amplification assay for viral detection. Nat Commun. 2020; 11(1):5920. PMC: 7679446. DOI: 10.1038/s41467-020-19258-y. View

5.
Mancuso C, Lu Z, Qian J, Boswell S, Springer M . A Semi-Quantitative Isothermal Diagnostic Assay Utilizing Competitive Amplification. Anal Chem. 2021; 93(27):9541-9548. PMC: 9837715. DOI: 10.1021/acs.analchem.1c01576. View