Xenon/computed Tomography Cerebral Blood Flow Measurements. Methods and Accuracy

Overview
Journal Invest Radiol
Specialty Radiology
Date 1987 Sep 1
PMID 3679761
Citations 4
Authors
Affiliations
Soon will be listed here.
Abstract

We performed a series of five baboon experiments to compare cerebral blood flow measured with an improved stable xenon/CT method and the radiolabelled microsphere technique at a PaCO2 of 40 mm Hg. The xenon/CT method was implemented by fitting the arterial xenon uptake with a double exponential function, by measuring the oxygen and carbon dioxide concentrations continuously during each breath and by taking into account the lung-to-brain transit time of xenon. The time of xenon inhalation was extended to 30 minutes to obtain more reliable estimates of CBF in white matter regions. The results indicate an overall correlation coefficient of 0.92 between the two methods and good numeric agreement.

Citing Articles

Ancillary Studies in Evaluating Pediatric Brain Death.

Henderson N, McDonald M J Pediatr Intensive Care. 2019; 6(4):234-239.

PMID: 31073456 PMC: 6260316. DOI: 10.1055/s-0037-1604015.


Persistent Metabolic Disturbance in the Perihemorrhagic Zone Despite a Normalized Cerebral Blood Flow Following Surgery for Intracerebral Hemorrhage.

Tobieson L, Rossitti S, Zsigmond P, Hillman J, Marklund N Neurosurgery. 2018; 84(6):1269-1279.

PMID: 29788388 PMC: 6520101. DOI: 10.1093/neuros/nyy179.


Physiologic Effects of Xenon in Xenon-CT Cerebral Blood Flow Studies on Comatose Patients.

Kosty J, Kofke W, Maloney-Wilensky E, Frangos S, Levine J, Leroux P Transl Stroke Res. 2013; 3(3):375-80.

PMID: 24323813 DOI: 10.1007/s12975-012-0183-8.


Simultaneous measurement of regional cerebral blood flow by perfusion CT and stable xenon CT: a validation study.

Wintermark M, Thiran J, Maeder P, Schnyder P, Meuli R AJNR Am J Neuroradiol. 2001; 22(5):905-14.

PMID: 11337336 PMC: 8174953.