» Articles » PMID: 36769537

Metal-on-Metal Hips: Ten-Year Clinical and Radiographic Outcomes of the ADEPT Metal-on-Metal Hip Resurfacing and Modular Total Hip Arthroplasty

Overview
Journal J Clin Med
Specialty General Medicine
Date 2023 Feb 11
PMID 36769537
Authors
Affiliations
Soon will be listed here.
Abstract

Background: The aim of this study is to update the 10-year follow-up survivorship and metal ions levels of a cohort of metal-on-metal (MoM) hip resurfacing (HR) and large-diameter-head (LDH) total hip arthroplasty (THA).

Methods: The study is a retrospective analysis of prospectively collected data that compared the outcomes of 24 MoM HR (21 patients) and 15 (11 patients) modular LHD MoM THA at >10 years follow-up. Baseline characteristics as well as intraoperative and postoperative information were collected, including complications, revisions, clinical and radiographic outcomes, and serum metal ions level (Cobalt, Chromium). Metal ion levels were compared using a two-tailed unpaired t-test and Wilcoxon signed-rank test (jamovi v2.3.3.0, Sydney, NSW, AU).

Results: No significant differences were detected in gender, BMI, and ASA score between the two groups. Patients in the modular THA group were significantly older (57 years vs. 46 years; < 0.05). The HR overall survivorship was 91.7% (22 of 24 hips) with survivorship from implant failure and/or aseptic loosening and/or metal debris related 100% of problems. The modular THA overall survivorship was 86.7% (13 of 15 hips) with survivorship from implant aseptic loosening and metal ions complications of 93.4% (14 of 15 hips). No significant difference was noted when comparing clinical outcomes. Metal ions were significantly lower in the HR group (Co 25.8 nmol/L vs. 89 nmol/L; < 0.001-Cr 33.5 nmol/L vs. 55.2 nmol/L; = 0.026).

Conclusion: Both implants reported excellent and comparable clinical outcomes at >10 years follow-up. The Adept HR reported remarkable survivorship, in line with the registry data, proving once again its reliability in young active males. The modular LDH THA, despite being discontinued, presented higher reliability and a lower failure rate when compared with similar withdrawn MoM implants. Trunnionosis did not appear to be a significant problem in this particular modular design.

Citing Articles

Breaking Barriers: Addressing Gender Disparities in Hip Resurfacing Surgery Access in the United States.

Parsa A, Ghorbani M, Mirzaei N, Ebrahimzadeh M Arch Bone Jt Surg. 2025; 12(11):810-812.

PMID: 39850922 PMC: 11756538. DOI: 10.22038/ABJS.2024.80526.3679.


Metallosis, Trunnionosis, and Head-stem Dissociation of the Exeter v40 Stem in a Metal-on-Metal Hip: A Unique Case.

Shet S, Mohan K, Gul R J Orthop Case Rep. 2025; 15(1):116-121.

PMID: 39801890 PMC: 11723720. DOI: 10.13107/jocr.2025.v15.i01.5146.


Metal-on-metal hip resurfacing arthroplasty: is it safe and reliable? A synopsis of the past, the present, and the future of HRA.

Xie F, Sheng S, Palan J, Pandit H EFORT Open Rev. 2024; 9(8):751-761.

PMID: 39087509 PMC: 11370715. DOI: 10.1530/EOR-23-0099.


Towards the First Generation of Biomimetic Fixation for Resurfacing Arthroplasty Endoprostheses.

Uklejewski R, Winiecki M, Dabrowski M, Rogala P Biomimetics (Basel). 2024; 9(2).

PMID: 38392145 PMC: 10887084. DOI: 10.3390/biomimetics9020099.

References
1.
Engh C, Bobyn J, Glassman A . Porous-coated hip replacement. The factors governing bone ingrowth, stress shielding, and clinical results. J Bone Joint Surg Br. 1987; 69(1):45-55. DOI: 10.1302/0301-620X.69B1.3818732. View

2.
Morlock M, Dickinson E, Gunther K, Bunte D, Polster V . Head Taper Corrosion Causing Head Bottoming Out and Consecutive Gross Stem Taper Failure in Total Hip Arthroplasty. J Arthroplasty. 2018; 33(11):3581-3590. DOI: 10.1016/j.arth.2018.07.017. View

3.
DeLee J, CHARNLEY J . Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976; (121):20-32. View

4.
Malahias M, Mancino F, Gu A, Adriani M, Martino I, Boettner F . Acetabular impaction grafting with mesh for acetabular bone defects: a systematic review. Hip Int. 2020; 32(2):185-196. DOI: 10.1177/1120700020971851. View

5.
Ridon P, Putman S, Migaud H, Berton C, Pasquier G, Girard J . Long-term comparative study of large-diameter metal-on-metal bearings: Resurfacing versus total arthroplasty with large-diameter Durom™ bearing. Orthop Traumatol Surg Res. 2019; 105(5):943-948. DOI: 10.1016/j.otsr.2019.04.006. View