Rapid Distortion Correction Enables Accurate Magnetic Resonance Imaging-guided Real-time Adaptive Radiotherapy

Overview
Specialty Oncology
Date 2023 Jan 30
PMID 36713071
Authors
Affiliations
Soon will be listed here.
Abstract

Background And Purpose: Magnetic resonance imaging (MRI)-Linac systems combine simultaneous MRI with radiation delivery, allowing treatments to be guided by anatomically detailed, real-time images. However, MRI can be degraded by geometric distortions that cause uncertainty between imaged and actual anatomy. In this work, we develop and integrate a real-time distortion correction method that enables accurate real-time adaptive radiotherapy.

Materials And Methods: The method was based on the pre-treatment calculation of distortion and the rapid correction of intrafraction images. A motion phantom was set up in an MRI-Linac at isocentre ( ), the edge ( ) and just outside ( ) the imaging volume. The target was irradiated and tracked during real-time adaptive radiotherapy with and without the distortion correction. The geometric tracking error and latency were derived from the measurements of the beam and target positions in the EPID images.

Results: Without distortion correction, the mean geometric tracking error was 1.3 mm at and 3.1 mm at . When distortion correction was applied, the error was reduced to 1.0 mm at and 1.1 mm at . The corrected error was similar to an error of 0.9 mm at where the target was unaffected by distortion indicating that this method has accurately accounted for distortion during tracking. The latency was 319 ± 12 ms without distortion correction and 335 ± 34 ms with distortion correction.

Conclusions: We have demonstrated a real-time distortion correction method that maintains accurate radiation delivery to the target, even at treatment locations with large distortion.

Citing Articles

Modeling of artificial intelligence-based respiratory motion prediction in MRI-guided radiotherapy: a review.

Zhang X, Yan D, Xiao H, Zhong R Radiat Oncol. 2024; 19(1):140.

PMID: 39380013 PMC: 11463122. DOI: 10.1186/s13014-024-02532-4.


Characterization of spatial integrity with active and passive implants in a low-field magnetic resonance linear accelerator scanner.

Pouymayou B, Perez-Haas Y, Allemann F, Saguner A, Andratschke N, Guckenberger M Phys Imaging Radiat Oncol. 2024; 30:100576.

PMID: 38644933 PMC: 11031795. DOI: 10.1016/j.phro.2024.100576.


Distortion-corrected image reconstruction with deep learning on an MRI-Linac.

Shan S, Gao Y, Liu P, Whelan B, Sun H, Dong B Magn Reson Med. 2023; 90(3):963-977.

PMID: 37125656 PMC: 10860740. DOI: 10.1002/mrm.29684.

References
1.
Caillet V, Zwan B, Briggs A, Hardcastle N, Szymura K, Prodreka A . Geometric uncertainty analysis of MLC tracking for lung SABR. Phys Med Biol. 2020; 65(23):235040. DOI: 10.1088/1361-6560/abb0c6. View

2.
Keijnemans K, Borman P, Uijtewaal P, Woodhead P, Raaymakers B, Fast M . A hybrid 2D/4D-MRI methodology using simultaneous multislice imaging for radiotherapy guidance. Med Phys. 2022; 49(9):6068-6081. PMC: 9545880. DOI: 10.1002/mp.15802. View

3.
Liu P, Dong B, Nguyen D, Ge Y, Hewson E, Waddington D . First experimental investigation of simultaneously tracking two independently moving targets on an MRI-linac using real-time MRI and MLC tracking. Med Phys. 2020; 47(12):6440-6449. DOI: 10.1002/mp.14536. View

4.
Baldwin L, Wachowicz K, Thomas S, Rivest R, Fallone B . Characterization, prediction, and correction of geometric distortion in 3 T MR images. Med Phys. 2007; 34(2):388-99. DOI: 10.1118/1.2402331. View

5.
Glitzner M, Woodhead P, Borman P, Lagendijk J, Raaymakers B . Technical note: MLC-tracking performance on the Elekta unity MRI-linac. Phys Med Biol. 2019; 64(15):15NT02. DOI: 10.1088/1361-6560/ab2667. View