Strengthening Fetal Heart Rate Monitoring During Labor with a Standard Handheld Doppler - Implementation Experience from India

Overview
Date 2022 Nov 28
PMID 36438535
Authors
Affiliations
Soon will be listed here.
Abstract

Background: India's neonatal and perinatal mortality is among the highest in the world. Intrapartum-related conditions contribute to a significant proportion of neonatal deaths and stillbirths. Fetal heart rate monitoring, a recommended norm to assess fetal well-bring, is not practiced as per standard guidelines in public health facilities. A standardized Doppler along with training on fetal heart rate monitoring was implemented across different levels of healthcare in three states.

Methods: Facilities were selected purposively to implement the Doppler. Baseline data for 3 months were collected. Interviews of health providers and observation of labor were conducted quarterly. Data were analyzed through a comparison of baseline and intervention on a number of delivery and monitoring indicators.

Results: Among 22,579 total deliveries, monitoring frequency increased along with increase in detection of abnormal fetal heart rate (FHR) while cesarean section and stillbirths reduced slightly. Cases never monitored reduced in the District Hospitals (7.98-2.07, < 0.01) and in Community Health Centers (14.7-1.67, < 0.001). Stillbirth rate reduced at the medical college (3.6-1.1, < 0.001). Interviews with providers revealed acceptance of the device due to its reliable readings.

Conclusion: The Doppler demonstrates acceptability and serves as a useful aid to improve intrapartum FHR monitoring.

References
1.
Herbst A, Ingemarsson I . Intermittent versus continuous electronic monitoring in labour: a randomised study. Br J Obstet Gynaecol. 1994; 101(8):663-8. DOI: 10.1111/j.1471-0528.1994.tb13181.x. View

2.
Ersdal H, Mduma E, Svensen E, Sundby J, Perlman J . Intermittent detection of fetal heart rate abnormalities identify infants at greatest risk for fresh stillbirths, birth asphyxia, neonatal resuscitation, and early neonatal deaths in a limited-resource setting: a prospective descriptive.... Neonatology. 2012; 102(3):235-42. DOI: 10.1159/000339481. View

3.
Sankar M, Neogi S, Sharma J, Chauhan M, Srivastava R, Prabhakar P . State of newborn health in India. J Perinatol. 2016; 36(s3):S3-S8. PMC: 5144119. DOI: 10.1038/jp.2016.183. View

4.
Lafontan S, Kidanto H, Ersdal H, Mbekenga C, Sundby J . Perceptions and experiences of skilled birth attendants on using a newly developed strap-on electronic fetal heart rate monitor in Tanzania. BMC Pregnancy Childbirth. 2019; 19(1):165. PMC: 6511185. DOI: 10.1186/s12884-019-2286-7. View

5.
McCusker J, Harris D, Hosmer Jr D . Association of electronic fetal monitoring during labor with cesarean section rate and with neonatal morbidity and mortality. Am J Public Health. 1988; 78(9):1170-4. PMC: 1349387. DOI: 10.2105/ajph.78.9.1170. View