Purpose: This study aims to determine whether changing the stem coating grants superior outcomes at a minimum follow-up of five years.

Methods: Retrospective review of a consecutive series of primary total hip arthroplasties (THAs) operated by direct anterior approach between 01/01/2013 and 31/12/2014. Two stems were compared, which were identical except for their surface coating; "the Original stem" was fully coated with hydroxyapatite (HA), while "the ProxCoat stem" was proximally coated with plasma-sprayed titanium and HA. Matching was performed. Clinical assessment included modified Harris hip score (mHHS), Oxford hip score (OHS), and forgotten joint score (FJS). Radiographic assessment evaluated alignment, subsidence, pedestal formation, heterotopic ossification, radiolucent lines ≥ 2 mm, spot welds, cortical hypertrophy, and osteolysis.

Results: 232 hips received the Original stem and 167 the ProxCoat stem, from which respectively five hips (2.2%) and no hips (0%) underwent revision. Matching identified two groups of 91 patients, with comparable patient demographics. At > five years follow-up, there were no differences in OHS (16 ± 6 vs 15 ± 5; p = 0.075) nor FJS (81 ± 26 vs 84 ± 22; p = 0.521), but there were differences in mHHS (89 ± 15 vs 92 ± 12; p = 0.042). There were no differences in alignment, subsidence, pedestal formation, heterotopic ossification, cortical hypertrophy, and osteolysis. There were differences in prevalence of proximal radiolucent lines (12% vs 0%; p < 0.001) and distal spot welds (24% vs 54%; p < 0.001).

Conclusion: At a minimum follow-up of five years, this study on matched patients undergoing primary THA found that ProxCoat stems results in significantly fewer radiolucent lines, more spot welds, and less revisions than Original stems, thus suggesting better bone ingrowth.

Citing Articles

High subsidence rate in primary total hip arthroplasty with a taper wedge stem featuring a three-dimensionally printed porous structure.

Ishikura H, Masuyama Y, Fujita S, Nishiwaki T Arch Orthop Trauma Surg. 2025; 145(1):188.

PMID: 40085230 DOI: 10.1007/s00402-025-05809-9.


Influence of proximal femoral morphology on osteointegration of the AMIS uncemented femoral stem in modern hip arthroplasty.

Maton M, Vandromme E, Charles T, Kapanci B, Jayankura M Arthroplasty. 2024; 6(1):56.

PMID: 39394176 PMC: 11470665. DOI: 10.1186/s42836-024-00274-y.


No differences in clinical and radiographic outcomes between standard offset and high offset short cementless stems.

DAmbrosio A, Bonnomet F, Debordes P, Ramos-Pascual S, Kumble A, Saffarini M Arch Orthop Trauma Surg. 2024; 144(8):3401-3411.

PMID: 39168958 DOI: 10.1007/s00402-024-05431-1.


No differences in clinical and radiographic outcomes between standard versus high offset collared stems for primary total hip arthroplasty at five years follow-up.

Jacquot L, Machenaud A, Grew B, Ramos-Pascual S, Dubreuil S, Saffarini M Int Orthop. 2024; 48(10):2525-2533.

PMID: 39112838 DOI: 10.1007/s00264-024-06264-8.


Improved radiographic outcomes and decreased complications rate on the femoral side can be achieved by a novel designed whole-process robotic assisted hip system for total hip arthroplasty: a prospective randomized controlled trial.

Huang Z, Zhang Z, Wang W, Wang G, Lu X, Zhang H J Robot Surg. 2024; 18(1):79.

PMID: 38365906 DOI: 10.1007/s11701-024-01835-8.