Peptide mapping by liquid chromatography mass spectrometry (LC-MS) and the related multi-attribute method (MAM) are well-established analytical tools for verification of the primary structure and mapping/quantitation of co- and post-translational modifications (PTMs) or product quality attributes in biopharmaceutical development. Proteolytic digestion is a key step in peptide mapping workflows, which traditionally is labor-intensive, involving multiple manual steps. Recently, simple high-temperature workflows with automatic digestion were introduced, which facilitate robustness and reproducibility across laboratories. Here, a modified workflow with an automatic digestion step is presented, which includes a two-step digestion at high and low temperatures, as opposed to the original one-step digestion at a high temperature. The new automatic digestion workflow significantly reduces the number of missed cleavages, obtaining a more complete digestion profile. In addition, we describe how chromatographic peak tailing and carry-over is dramatically reduced for hydrophobic peptides by switching from the traditional C18 reversed-phase (RP) column chemistry used for peptide mapping to a less retentive C4 column chemistry. No negative impact is observed on MS/MS-derived sequence coverage when switching to a C4 column chemistry. Overall, the new peptide mapping workflow significantly reduces the number of missed cleavages, yielding more robust and simple data interpretation, while providing dramatically reduced tailing and carry-over of hydrophobic peptides.

Citing Articles

Clean and Complete Protein Digestion with an Autolysis Resistant Trypsin for Peptide Mapping.

Muriithi B, Ippoliti S, Finny A, Addepalli B, Lauber M J Proteome Res. 2024; 23(11):5221-5228.

PMID: 39392678 PMC: 11536465. DOI: 10.1021/acs.jproteome.4c00598.


Addressing Acid-Catalyzed Deamidation and the Solubility of Hydrophobic Peptides in Multi-Attribute Method Workflows.

Kristensen D, Orgaard M, Sloth T, Comamala G, Jensen P Anal Chem. 2023; 95(42):15465-15471.

PMID: 37824441 PMC: 10603607. DOI: 10.1021/acs.analchem.3c02609.


Mass spectrometry-based multi-attribute method in protein therapeutics product quality monitoring and quality control.

Yang F, Zhang J, Buettner A, Vosika E, Sadek M, Hao Z MAbs. 2023; 15(1):2197668.

PMID: 37057828 PMC: 10114992. DOI: 10.1080/19420862.2023.2197668.


Interlaboratory Evaluation of a User-Friendly Benchtop Mass Spectrometer for Multiple-Attribute Monitoring Studies of a Monoclonal Antibody.

Butre C, DAtri V, Diemer H, Colas O, Wagner E, Beck A Molecules. 2023; 28(6).

PMID: 36985827 PMC: 10053224. DOI: 10.3390/molecules28062855.

References
1.
Ren D . Advancing Mass Spectrometry Technology in cGMP Environments. Trends Biotechnol. 2020; 38(10):1051-1053. DOI: 10.1016/j.tibtech.2020.06.007. View

2.
Rogstad S, Faustino A, Ruth A, Keire D, Boyne M, Park J . A Retrospective Evaluation of the Use of Mass Spectrometry in FDA Biologics License Applications. J Am Soc Mass Spectrom. 2016; 28(5):786-794. DOI: 10.1007/s13361-016-1531-9. View

3.
Aqvist J, Socan J, Purg M . Hidden Conformational States and Strange Temperature Optima in Enzyme Catalysis. Biochemistry. 2020; 59(40):3844-3855. PMC: 7584337. DOI: 10.1021/acs.biochem.0c00705. View

4.
Rogstad S, Yan H, Wang X, Powers D, Brorson K, Damdinsuren B . Multi-Attribute Method for Quality Control of Therapeutic Proteins. Anal Chem. 2019; 91(22):14170-14177. DOI: 10.1021/acs.analchem.9b03808. View

5.
Ogata Y, Quizon P, Nightlinger N, Sitasuwan P, Snodgrass C, Lee L . Automated multi-attribute method sample preparation using high-throughput buffer exchange tips. Rapid Commun Mass Spectrom. 2021; 36(3):e9222. PMC: 9286584. DOI: 10.1002/rcm.9222. View