» Articles » PMID: 36320718

Balloon-Expandable TAVR Bioprostheses: Area or Perimeter Sizing? A Prospective Pilot Study

Overview
Publisher Wiley
Date 2022 Nov 2
PMID 36320718
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: In TAVR, area sizing is used for balloon-expandable (BE) valves, whereas self-expanding valves are sized to annulus perimeter. For BE valves, this seems illogical: these frames force a circular shape even on an ellipsoid annulus. This can potentially lead to relative undersizing when area sizing is being applied. We developed a perimeter-based sizing algorithm to evaluate the safety and feasibility of perimeter sizing for the Myval BE valve.

Methods: In this prospective single-center study, 60 patients with severe aortic stenosis treated with the Myval BE valve were included. Perimeter sizing was used with limited oversizing of 3.7% ± 1.3% compared to the annulus perimeter. After TAVR, clinical outcomes were evaluated at 30 days and 1 year. An echocardiographic follow-up took place at 30 days.

Results: At 30 days, the need for PPI and stroke occurred in 2% and 3% of the patients, respectively. Moreover, cardiac death and moderate-severe PVL were absent. At 1-year, cardiac death and stroke were observed in 3% and 8% of the patients, respectively. In 33.3% of the patients, a larger valve size was implanted compared to the valve size calculated by area sizing.

Conclusions: Perimeter sizing with the Myval BE valve leads to substantial use of larger valve sizes and favorable clinical outcomes, with low PPI and the absence of significant PVL. A randomized controlled trial is being planned to prove the superiority of this alternative sizing method.

Citing Articles

Myval Transcatheter Heart Valve: The Future of Transcatheter Valve Replacement and Significance in Current Timeline.

Kilic T, Coskun S, Mirzamidinov D, Yilmaz I, Yavuz S, Sahin T J Clin Med. 2024; 13(22).

PMID: 39598000 PMC: 11594825. DOI: 10.3390/jcm13226857.

References
1.
Leon M, Smith C, Mack M, Makkar R, Svensson L, Kodali S . Transcatheter or Surgical Aortic-Valve Replacement in Intermediate-Risk Patients. N Engl J Med. 2016; 374(17):1609-20. DOI: 10.1056/NEJMoa1514616. View

2.
Mack M, Leon M, Thourani V, Makkar R, Kodali S, Russo M . Transcatheter Aortic-Valve Replacement with a Balloon-Expandable Valve in Low-Risk Patients. N Engl J Med. 2019; 380(18):1695-1705. DOI: 10.1056/NEJMoa1814052. View

3.
Gurvitch R, Tay E, Wijesinghe N, Ye J, Nietlispach F, Wood D . Transcatheter aortic valve implantation: lessons from the learning curve of the first 270 high-risk patients. Catheter Cardiovasc Interv. 2011; 78(7):977-84. DOI: 10.1002/ccd.22961. View

4.
Mehrotra P, Flynn A, Jansen K, Tan T, Mak G, Julien H . Differential left ventricular outflow tract remodeling and dynamics in aortic stenosis. J Am Soc Echocardiogr. 2015; 28(11):1259-66. DOI: 10.1016/j.echo.2015.07.018. View

5.
Barbanti M, Yang T, Cabau J, Tamburino C, Wood D, Jilaihawi H . Anatomical and procedural features associated with aortic root rupture during balloon-expandable transcatheter aortic valve replacement. Circulation. 2013; 128(3):244-53. DOI: 10.1161/CIRCULATIONAHA.113.002947. View