Abstract

Background: Rapid diagnostic tests (RDTs) are being widely used to manage COVID-19 pandemic. However, many results remain unreported or unconfirmed, altering a correct epidemiological surveillance.

Objective: Our aim was to evaluate an artificial intelligence-based smartphone app, connected to a cloud web platform, to automatically and objectively read RDT results and assess its impact on COVID-19 pandemic management.

Methods: Overall, 252 human sera were used to inoculate a total of 1165 RDTs for training and validation purposes. We then conducted two field studies to assess the performance on real-world scenarios by testing 172 antibody RDTs at two nursing homes and 96 antigen RDTs at one hospital emergency department.

Results: Field studies demonstrated high levels of sensitivity (100%) and specificity (94.4%, CI 92.8%-96.1%) for reading IgG band of COVID-19 antibody RDTs compared to visual readings from health workers. Sensitivity of detecting IgM test bands was 100%, and specificity was 95.8% (CI 94.3%-97.3%). All COVID-19 antigen RDTs were correctly read by the app.

Conclusions: The proposed reading system is automatic, reducing variability and uncertainty associated with RDTs interpretation and can be used to read different RDT brands. The web platform serves as a real-time epidemiological tracking tool and facilitates reporting of positive RDTs to relevant health authorities.

Citing Articles

Smartphone-based device for point-of-care diagnostics of pulmonary inflammation using convolutional neural networks (CNNs).

Ghaderinia M, Abadijoo H, Mahdavian A, Kousha E, Shakibi R, Taheri S Sci Rep. 2024; 14(1):6912.

PMID: 38519489 PMC: 10959990. DOI: 10.1038/s41598-024-54939-4.


Measuring the performance of computer vision artificial intelligence to interpret images of HIV self-testing results.

Roche S, Ekwunife O, Mendonca R, Kwach B, Omollo V, Zhang S Front Public Health. 2024; 12:1334881.

PMID: 38384878 PMC: 10880864. DOI: 10.3389/fpubh.2024.1334881.


Digital Platform for Automatic Qualitative and Quantitative Reading of a Cryptococcal Antigen Point-of-Care Assay Leveraging Smartphones and Artificial Intelligence.

Bermejo-Pelaez D, Medina N, Alamo E, Soto-Debran J, Bonilla O, Luengo-Oroz M J Fungi (Basel). 2023; 9(2).

PMID: 36836331 PMC: 9961444. DOI: 10.3390/jof9020217.

References
1.
Turbe V, Herbst C, Mngomezulu T, Meshkinfamfard S, Dlamini N, Mhlongo T . Deep learning of HIV field-based rapid tests. Nat Med. 2021; 27(7):1165-1170. PMC: 7611654. DOI: 10.1038/s41591-021-01384-9. View

2.
Mendels D, Dortet L, Emeraud C, Oueslati S, Girlich D, Ronat J . Using artificial intelligence to improve COVID-19 rapid diagnostic test result interpretation. Proc Natl Acad Sci U S A. 2021; 118(12). PMC: 7999948. DOI: 10.1073/pnas.2019893118. View

3.
Borges L, Martins A, de Mello Silva B, de Paula Dias B, Goncalves R, de Souza D . Rapid diagnosis of COVID-19 in the first year of the pandemic: A systematic review. Int Immunopharmacol. 2021; 101(Pt A):108144. PMC: 8440261. DOI: 10.1016/j.intimp.2021.108144. View

4.
Weber N, Klepser M, Akers J, Klepser D, Adams A . Use of CLIA-waived point-of-care tests for infectious diseases in community pharmacies in the United States. Expert Rev Mol Diagn. 2015; 16(2):253-64. DOI: 10.1586/14737159.2015.1116388. View

5.
Richardson S, Kohn M, Bollyky J, Parsonnet J . Validity of at-home rapid antigen lateral flow assay and artificial intelligence read to detect SARS-CoV-2. Diagn Microbiol Infect Dis. 2022; 104(3):115763. PMC: 9259514. DOI: 10.1016/j.diagmicrobio.2022.115763. View