Long-term Consequences of Ventricular Septal Defect Closure Using Nit-Occlud Le VSD Coil Device: A Systematic Review and Meta-Analysis

Overview
Journal Heart Views
Date 2022 Oct 10
PMID 36213425
Authors
Affiliations
Soon will be listed here.
Abstract

Background: The successful management of ventricular septal defect (VSD) has been possible through the development of advanced techniques. In this regard, percutaneous VSD closure by employing different types of occluders as an alternative for surgery can help to achieve the most desirable postprocedural consequences. However, the studies reported contradictory results on the use of different brands of VSD occluders. Herein, we performed a systematic review and meta-analysis of published studies to assess pooled long-term success rate and potential complications of using the Nit-Occlud Lê VSD coil for VSD closure.

Materials And Methods: Two reviewers began to deeply search the various databases for all eligible studies in accordance with the considered keywords. The inclusion criterion for retrieving the studies was to describe the mid-term or long-term consequences of VSD closing by the Nit-Occlud Lê VSD coil device. In the final, eight articles were eligible for the analysis. The follow-up time of the studies ranged from 6 months to 5 years.

Results: The success rate of the procedure ranged from 87.0% to 100% considering the weight of each study, the pooled success rate of VSDs closure by Nit-Occlud Lê VSD coil device was 93.1% (95% confidence interval [CI]: 89.9% to 95.5%). The pooled prevalence of postprocedural residual shunt was estimated to be 9.6% (95%CI: 6.8% to 13.4%). The corrected pooled prevalence of trivial mild aortic regurgitation (AR) was 2.9% (95%CI: 1.5% to 5.4%); however, moderate-to-severe AR and complete heart block or right bundle branch block were shown to be rare.

Conclusion: VSD closure using a Nit-Occlud Lê VSD coil device can lead to a high success rate with low rates of residual shunt, cardiac conductive or vascular disturbances.

Citing Articles

Experience of percutaneous closure of small to large patent ductus arteriosus with Nit-Occlud device in a tertiary referral hospital in Colombia.

Galvis-Luna S, Guzman-Serrano C, Olave-Rueda I, Mejia-Quinones V, Gutierrez-Gil J, Mosquera-Alvarez W Arch Cardiol Mex. 2024; 94(3):263-268.

PMID: 38467114 PMC: 11259415. DOI: 10.24875/ACM.23000267.

References
1.
Kang S, Tometzki A, Caputo M, Morgan G, Parry A, Martin R . Longer-term outcome of perventricular device closure of muscular ventricular septal defects in children. Catheter Cardiovasc Interv. 2015; 85(6):998-1005. DOI: 10.1002/ccd.25821. View

2.
Houeijeh A, Godart F, Jalal Z, Ovaert C, Heitz F, Mauran P . Transcatheter closure of a perimembranous ventricular septal defect with Nit-Occlud Lê VSD Coil: A French multicentre study. Arch Cardiovasc Dis. 2020; 113(2):104-112. DOI: 10.1016/j.acvd.2019.11.004. View

3.
Bu H, Yang Y, Hu S, Wu Q, Gong X, Zhao T . A novel biodegradable occluder for the closure of ventricular septal defects: immediate and medium-term results in a canine model. Interact Cardiovasc Thorac Surg. 2019; 29(5):783-792. DOI: 10.1093/icvts/ivz174. View

4.
Bulut M, Kucuk M, Balli S, Celebi A . Treatment of severe hemolysis following Nit-Occlud Lê VSD coil implantation with Amplatzer Duct Occluder II. Turk Kardiyol Dern Ars. 2016; 44(7):593-596. DOI: 10.5543/tkda.2016.69841. View

5.
Pedra C, Pontes Jr S, Pedra S, Salerno L, Sousa J, Miaira M . Percutaneous closure of postoperative and post-traumatic ventricular septal defects. J Invasive Cardiol. 2007; 19(11):491-5. View