Applicability of Devices Available for the Measurement of Intracompartmental Pressures: a Cadaver Study

Overview
Journal J Exp Orthop
Publisher Wiley
Specialty Orthopedics
Date 2022 Sep 27
PMID 36166161
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: The indication for surgical treatment of the chronic exertional compartment syndrome is evaluated by measuring intracompartmental pressures. The validity of these invasive intracompartmental pressure measurements are increasingly questioned in the absence of a standardized test protocol and uniform cut-off values. The aim of the current study was to test compartment pressure monitors and needles for uniformity, thereby supporting the physician's choice in the selection of appropriate test materials.

Methods: A compartment syndrome was simulated in embalmed above-knee cadaveric leg specimen. Four different terminal devices (Compass manometer, Stryker device, Meritrans transduce, and arterial line) were tested with 22 different needle types. Legs were pressurized after introduction of the four terminal devices in the anterior compartment, using the same needle type. Pressure was recorded at a 30-second interval for 11 minutes in total. Before and after pressurization, the intravenous bag of saline was weighed.

Results: The simulation of a compartment syndrome resulted in intracompartmental pressure values exceeding 100 mmHg in 17 of the 22 legs (77%). In the other five legs, a smaller built-up of pressure was seen, although maximum intracompartmental pressure was in between 70 and 100 mmHg. The intraclass correlation coefficient was above 0.700 for all possible needle types. Excellent to good resemblance was seen in 16 out of 22 instrumental setups (73%). The mean volume of saline infusion required in runs that exceeded 100 mmHg (309 ± 116 ml) was significantly lower compared to the legs in which 100 mmHg was not achieved (451 ± 148 ml; p = 0.04).

Conclusion: The intracompartmental pressure recordings of the four terminal devices were comparable, when tested with a standardized pressurization model in a human cadaver model. None of the included terminal devices or needle types were found to be superior. The results provide evidence for more diverse material selection when logistic choices for intracompartmental pressure measurement devices are warranted.

Level Of Evidence: Level IV.

Citing Articles

[Diagnostics of acute compartment syndrome : Current gold standard and the state of science of noninvasive assessment methods].

Sellei R, Kobbe P, Pape H, Hildebrand F Chirurgie (Heidelb). 2024; 95(7):529-538.

PMID: 38806712 DOI: 10.1007/s00104-024-02096-9.


Novel Insights Into the Intratendinous Pressure Behavior of the Achilles Tendon in Athletes.

Pringels L, Van Acker G, Wezenbeek E, Burssens A, Vanden Bossche L Sports Health. 2024; 17(2):374-382.

PMID: 38610105 PMC: 11569691. DOI: 10.1177/19417381241245357.

References
1.
Shuler F, Dietz M . Physicians' ability to manually detect isolated elevations in leg intracompartmental pressure. J Bone Joint Surg Am. 2010; 92(2):361-7. DOI: 10.2106/JBJS.I.00411. View

2.
Roberts A, Franklyn-Miller A . The validity of the diagnostic criteria used in chronic exertional compartment syndrome: a systematic review. Scand J Med Sci Sports. 2011; 22(5):585-95. DOI: 10.1111/j.1600-0838.2011.01386.x. View

3.
Vogels S, Ritchie E, Bakker E, Vogels M, Zimmermann W, Verhofstad M . Measuring intracompartmental pressures for the chronic exertional compartment syndrome: Challenging commercially available devices and their respective accuracy. J Biomech. 2022; 135:111026. DOI: 10.1016/j.jbiomech.2022.111026. View

4.
Teng A, Huang J, Wilber R, Wilber J . Treatment of compartment syndrome: transverse fasciotomy as an adjunct to longitudinal dermatofasciotomy: an in vitro study. J Orthop Trauma. 2005; 19(7):442-7. DOI: 10.1097/01.bot.0000161539.16656.08. View

5.
Gatt A, Schembri-Wismayer P, Chockalingam N, Formosa C . Kinematic and Kinetic Comparison of Fresh Frozen and Thiel-Embalmed Human Feet for Suitability for Biomechanical Educational and Research Settings. J Am Podiatr Med Assoc. 2019; 109(2):113-121. DOI: 10.7547/16-130. View