» Articles » PMID: 36132820

Nanoscale Heat Transport Analysis by Scanning Thermal Microscopy: from Calibration to High-resolution Measurements

Overview
Journal Nanoscale Adv
Specialty Biotechnology
Date 2022 Sep 22
PMID 36132820
Authors
Affiliations
Soon will be listed here.
Abstract

Scanning thermal microscopy (SThM) is a powerful technique for thermal characterization. However, one of the most challenging aspects of thermal characterization is obtaining quantitative information on thermal conductivity with nanoscale lateral resolution. We used this technique with the cross-point calibration method to obtain the thermal contact resistance, , and thermal exchange radius, , using thermo-resistive Pd/SiN probes. The cross-point curves correlate the dependence of and with the sample's thermal conductivity. We implemented a 3-SThM method in which reference samples with known thermal conductivity were used in the calibration and validation process to guarantee optimal working conditions. We achieved values of = 0.94 × 10 ± 0.02 K W and = 2.41 × 10 ± 0.02 m for samples with a low thermal conductivity (between 0.19 and 1.48 W m K). These results show a large improvement in spatial resolution over previously reported data for the Wollaston probes (where ∼ 2.8 μm). Furthermore, the contact resistance with the Pd/SiN is ∼20× larger than reported for a Wollaston wire probe (with 0.45 × 10 K W). These thermal parameters were used to determine the unknown thermal conductivity of thermoelectric films of AgSe, Ag Se, CuSe (smooth rough surface), and BiTe, obtaining, in units of W m K, the values of 0.63 ± 0.07, 0.69 ± 0.15, 0.79 ± 0.03, 0.82 ± 0.04, and 0.93 ± 0.12, respectively. To the best of our knowledge, this is the first time these microfabricated probes have been calibrated using the cross-point method to perform quantitative thermal analysis with nanoscale resolution. Moreover, this work shows high-resolution thermal images of the and signals, which can offer relevant information on the material's heat dissipation.

References
1.
Thiery L, Gavignet E, Cretin B . Two omega method for active thermocouple microscopy. Rev Sci Instrum. 2009; 80(3):034901. DOI: 10.1063/1.3097183. View

2.
de Lucas-Gil E, Reinosa J, Neuhaus K, Vera-Londono L, Martin-Gonzalez M, Fernandez J . Exploring New Mechanisms for Effective Antimicrobial Materials: Electric Contact-Killing Based on Multiple Schottky Barriers. ACS Appl Mater Interfaces. 2017; 9(31):26219-26225. DOI: 10.1021/acsami.7b09695. View

3.
Ge Y, Zhang Y, Booth J, Weaver J, Dobson P . Quantification of probe-sample interactions of a scanning thermal microscope using a nanofabricated calibration sample having programmable size. Nanotechnology. 2016; 27(32):325503. DOI: 10.1088/0957-4484/27/32/325503. View

4.
Munoz Rojo M, Abad B, Manzano C, Torres P, Cartoixa X, Alvarez F . Thermal conductivity of BiTe nanowires: how size affects phonon scattering. Nanoscale. 2017; 9(20):6741-6747. DOI: 10.1039/c7nr02173a. View

5.
Liu J, Zhu J, Tian M, Gu X, Schmidt A, Yang R . Simultaneous measurement of thermal conductivity and heat capacity of bulk and thin film materials using frequency-dependent transient thermoreflectance method. Rev Sci Instrum. 2013; 84(3):034902. DOI: 10.1063/1.4797479. View