Technical Design Considerations of a Human-Scale Talbot-Lau Interferometer for Dark-Field CT

Overview
Date 2022 Sep 16
PMID 36112565
Authors
Affiliations
Soon will be listed here.
Abstract

Computed tomography (CT) as an important clinical diagnostics method can profit from extension with dark-field imaging, as it is currently restricted to X-rays' attenuation contrast only. Dark-field imaging allows access to more tissue properties, such as micro-structural texture or porosity. The up-scaling process to clinical scale is complex because several design constraints must be considered. The two most important ones are that the finest grating is limited by current manufacturing technology to a [Formula: see text] period and that the interferometer should fit into the CT gantry with minimal modifications only. In this work we discuss why an inverse interferometer and a triangular G1 profile are advantageous and make a compact and sensitive interferometer implementation feasible. Our evaluation of the triangular grating profile reveals a deviation in the interference pattern compared to standard grating profiles, which must be considered in the subsequent data processing. An analysis of the grating orientation demonstrates that currently only a vertical layout can be combined with cylindrical bending of the gratings. We also provide an in-depth discussion, including a new simulation approach, of the impact of the extended X-ray source spot which can lead to large performance loss and present supporting experimental results. This analysis reveals a vastly increased sensitivity to geometry and grating period deviations, which must be considered early in the system design process.

Citing Articles

[Dark-field imaging and computed tomography : Novel X-ray-based contrast imaging modality with great promise for pulmonary imaging].

Pfeiffer F, Willer K, Viermetz M, Pfeiffer D Radiologie (Heidelb). 2023; 63(7):513-522.

PMID: 37341743 DOI: 10.1007/s00117-023-01161-4.


The choice of an autocorrelation length in dark-field lung imaging.

Spindler S, Etter D, Rawlik M, Polikarpov M, Romano L, Shi Z Sci Rep. 2023; 13(1):2731.

PMID: 36792717 PMC: 9932147. DOI: 10.1038/s41598-023-29762-y.


Dark-field chest X-ray imaging for the assessment of COVID-19-pneumonia.

Frank M, Gassert F, Urban T, Willer K, Noichl W, Schick R Commun Med (Lond). 2022; 2(1):147.

PMID: 36411311 PMC: 9678896. DOI: 10.1038/s43856-022-00215-3.