» Articles » PMID: 36040972

Detailed Measurements and Simulations of Electric Field Distribution of Two TMS Coils Cleared for Obsessive Compulsive Disorder in the Brain and in Specific Regions Associated with OCD

Overview
Journal PLoS One
Date 2022 Aug 30
PMID 36040972
Authors
Affiliations
Soon will be listed here.
Abstract

The FDA cleared deep transcranial magnetic stimulation (Deep TMS) with the H7 coil for obsessive-compulsive disorder (OCD) treatment, following a double-blinded placebo-controlled multicenter trial. Two years later the FDA cleared TMS with the D-B80 coil on the basis of substantial equivalence. In order to investigate the induced electric field characteristics of the two coils, these were placed at the treatment position for OCD over the prefrontal cortex of a head phantom, and the field distribution was measured. Additionally, numerical simulations were performed in eight Population Head Model repository models with two sets of conductivity values and three Virtual Population anatomical head models and their homogeneous versions. The H7 was found to induce significantly higher maximal electric fields (p<0.0001, t = 11.08) and to stimulate two to five times larger volumes in the brain (p<0.0001, t = 6.71). The rate of decay of electric field with distance is significantly slower for the H7 coil (p < 0.0001, Wilcoxon matched-pairs test). The field at the scalp is 306% of the field at a 3 cm depth with the D-B80, and 155% with the H7 coil. The H7 induces significantly higher intensities in broader volumes within the brain and in specific brain regions known to be implicated in OCD (dorsal anterior cingulate cortex (dACC), dorsolateral prefrontal cortex (dlPFC), inferior frontal gyrus (IFG), orbitofrontal cortex (OFC) and pre-supplementary motor area (pre-SMA)) compared to the D-B80. Significant field ≥ 80 V/m is induced by the H7 (D-B80) in 15% (1%) of the dACC, 78% (29%) of the pre-SMA, 50% (20%) of the dlPFC, 30% (12%) of the OFC and 15% (1%) of the IFG. Considering the substantial differences between the two coils, the clinical efficacy in OCD should be tested and verified separately for each coil.

Citing Articles

Trends of socioeconomic and geographic inequalities in COVID-19 vaccine uptake in Ethiopia: using the WHO Health Equity Assessment Toolkit.

Endawkie A, Daba C, Asmare L, Desye B, Mawugatie T, Melak D BMC Health Serv Res. 2024; 24(1):1629.

PMID: 39707307 PMC: 11660808. DOI: 10.1186/s12913-024-12082-w.


Ultra-high frequency repetitive TMS at subthreshold intensity induces suprathreshold motor response via temporal summation.

Nguyen H, Li C, Hoffman S, Deng Z, Yang Y, Lu H J Neural Eng. 2024; 21(4).

PMID: 39079555 PMC: 11307324. DOI: 10.1088/1741-2552/ad692f.


Outcome measures for electric field modeling in tES and TMS: A systematic review and large-scale modeling study.

Hoornweder S, Nuyts M, Frieske J, Verstraelen S, Meesen R, Caulfield K Neuroimage. 2023; 281:120379.

PMID: 37716590 PMC: 11008458. DOI: 10.1016/j.neuroimage.2023.120379.


Will Transcranial Magnetic Stimulation Improve the Treatment of Obsessive-Compulsive Disorder? A Systematic Review and Meta-Analysis of Current Targets and Clinical Evidence.

Grassi G, Moradei C, Cecchelli C Life (Basel). 2023; 13(7).

PMID: 37511869 PMC: 10381766. DOI: 10.3390/life13071494.

References
1.
Abramowitz J, Taylor S, McKay D . Obsessive-compulsive disorder. Lancet. 2009; 374(9688):491-9. DOI: 10.1016/S0140-6736(09)60240-3. View

2.
Tzirini M, Roth Y, Harmelech T, Zibman S, Pell G, Kimiskidis V . Electrical field measurements and simulations of the H7 and D-B80 coils: Non-equivalence of the TMS coils for obsessive compulsive disorder. Brain Stimul. 2021; 14(6):1525-1527. DOI: 10.1016/j.brs.2021.10.382. View

3.
Roth Y, Zangen A, Hallett M . A coil design for transcranial magnetic stimulation of deep brain regions. J Clin Neurophysiol. 2002; 19(4):361-70. DOI: 10.1097/00004691-200208000-00008. View

4.
Modirrousta M, Meek B, Sareen J, Enns M . Impaired trial-by-trial adjustment of cognitive control in obsessive compulsive disorder improves after deep repetitive transcranial magnetic stimulation. BMC Neurosci. 2015; 16:63. PMC: 4599438. DOI: 10.1186/s12868-015-0205-z. View

5.
Schmidtke K, Schorb A, Winkelmann G, Hohagen F . Cognitive frontal lobe dysfunction in obsessive-compulsive disorder. Biol Psychiatry. 1998; 43(9):666-73. DOI: 10.1016/s0006-3223(97)00355-7. View