» Articles » PMID: 36033225

A Multi-center Trial on Efficacy and Safety of the LifeTech CeraFlex ASD Occluder for Transcatheter Closure in Patients with Secundum Atrial Septal Defects

Overview
Authors
Affiliations
Soon will be listed here.
Abstract

Background: The last decades have brought remarkable improvements in treatment strategy and occluder modification of secundum atrial septal defect (ASD) closure. Approval, efficacy and safety of ASD closure devices have previously been demonstrated. This study investigated the clinical efficacy and safety of the LifeTech CeraFlex ASD occluder for interventional closure of secundum ASD with a 6-month follow-up (FU).

Methods: Procedure specific data was collected on patients considered for ASD closure with the CeraFlex occluder between April 2016 and December 2019 in three German centers. Efficacy and safety were assessed after device closure, at discharge, and at 6-month FU.

Results: The primary endpoint (successful ASD closure without severe complications) was reached by 102/103 patients (99%). Device embolization occurred in two patients (one early and one late embolization). After early snare-retrieval of an embolized device, this ASD was closed surgically and in the other patient with late device embolization the defect was closed with a larger CeraFlex occluder. The secondary endpoint (clincal efficacy after 6 months) was reached by 94/98 patients since new onset of arrhythmia occurred in four patients. Three patients had withdrawn their study-participation and one patient had moderate residual shunt, but not related to the occluder. Incomplete right bundle branch block (iRBBB) was seen in 31 patients. At last FU only 17 patients had remaining iRBBB documenting effective volume unloading of the right ventricle.

Conclusions: Catheter interventional closure of secundum ASDs with the CeraFlex ASD occluder was feasible, safe and effective in this study.

Citing Articles

Arrhythmias Following Patent Foramen Ovale Closure: An Unsolved Enigma.

Karanikola A, Soulaidopoulos S, Leontsinis I, Dri E, Sagris M, Kordalis A Life (Basel). 2025; 14(12.

PMID: 39768297 PMC: 11678317. DOI: 10.3390/life14121590.


Pulmonary hypertension in adults with congenital heart defects (ACHDs) in light of the 2022 ESC PAH guidelines-part II: supportive therapy, special situations (pregnancy, contraception, non-cardiac surgery), targeted pharmacotherapy, organ....

Kaemmerer H, Diller G, Dahnert I, Achenbach S, Eichstaedt C, Eicken A Cardiovasc Diagn Ther. 2024; 14(5):921-934.

PMID: 39513142 PMC: 11538840. DOI: 10.21037/cdt-24-167.


Recent Development of Biodegradable Occlusion Devices for Intra-Atrial Shunts.

Li Y, Chen Z, Xie Z, Wang S, Xie Y, Zhang Z Rev Cardiovasc Med. 2024; 25(5):159.

PMID: 39076475 PMC: 11267192. DOI: 10.31083/j.rcm2505159.


Transcatheter Closure of Secundum Atrial Septal Defects in Pediatric Patients: A 15-Year Single-Center Experience.

Dogan E, Levent E Cureus. 2024; 16(3):e57150.

PMID: 38681362 PMC: 11055965. DOI: 10.7759/cureus.57150.


Nanotechnology Innovations in Pediatric Cardiology and Cardiovascular Medicine: A Comprehensive Review.

Moisa S, Burlacu A, Butnariu L, Vasile C, Brinza C, Spoiala E Biomedicines. 2024; 12(1).

PMID: 38255290 PMC: 10813221. DOI: 10.3390/biomedicines12010185.


References
1.
Vincent R, Raviele A, Diehl H . Single-center experience with the HELEX septal occluder for closure of atrial septal defects in children. J Interv Cardiol. 2003; 16(1):79-82. DOI: 10.1046/j.1540-8183.2003.08005.x. View

2.
Cowley C, Lloyd T, Bove E, Gaffney D, Dietrich M, Rocchini A . Comparison of results of closure of secundum atrial septal defect by surgery versus Amplatzer septal occluder. Am J Cardiol. 2001; 88(5):589-91. DOI: 10.1016/s0002-9149(01)01750-7. View

3.
Chen T, Hsiao Y, Cheng C, Mao C, Chen D, Tsai M . In-Hospital and 4-Year Clinical Outcomes Following Transcatheter Versus Surgical Closure for Secundum Atrial Septal Defect in Adults: A National Cohort Propensity Score Analysis. Medicine (Baltimore). 2015; 94(38):e1524. PMC: 4635747. DOI: 10.1097/MD.0000000000001524. View

4.
Moore J, Hegde S, El-Said H, Beekman 3rd R, Benson L, Bergersen L . Transcatheter device closure of atrial septal defects: a safety review. JACC Cardiovasc Interv. 2013; 6(5):433-42. DOI: 10.1016/j.jcin.2013.02.005. View

5.
Haas N, Happel C, Soetemann D, Hanslik A, Moysich A, Kececioglu D . Optimal septum alignment of the Figulla Flex occluder to the atrial septum in patients with secundum atrial septal defects. EuroIntervention. 2014; 11(10):1153-60. DOI: 10.4244/EIJY14M12_09. View