Analytical Validation of a Portable Human Accu-Chek Glucometer in Honeybee Hemolymph

Overview
Date 2022 Aug 22
PMID 35993265
Authors
Affiliations
Soon will be listed here.
Abstract

Glucose and trehalose are the main energy sources used by honeybees () for daily activities. However, there is no validated point-of-care method to reliably measure both sugars. We performed an analytical validation of a portable human glucometer (Accu-Chek; Roche) for glucose measurement in honeybee hemolymph compared to a reference method (GluCH, UniCel DxC 600; Beckman Coulter). We used 30 pooled hemolymph samples collected from the antennae of anesthetized honeybees and diluted 1:4 in 0.9% saline. We evaluated dilution linearity, spike recovery, and inter- and intra-assay imprecision. Glucose concentration was measured over time (2 h, 4 h, 8 h, 12 h, 1 d, 2 d, 3 d, 7 d, 21 d, 28 d) at various storage temperature (25°C, 4°C, -20°C, -80°C). The trehalose concentration was measured indirectly by trehalase hydrolyzation. Glucose concentrations measured by both instruments had a strong correlation (0.985,  < 0.0001) and a bias of -7.33 mmol/L (±1.96SD: 13.70 to -28.36), with linear agreement at <20 mmol/L (physiologic value: 100 mmol/L). The accuracy of the glucometer decreased at >20 mmol/L. Recovery of 115-130% of diluted spikes indicated good specificity. Inter- and intra-assay imprecision were 2.50% and 2.21%, respectively. Glucose concentrations fluctuated in stored samples dependent on time and temperature; however, glucose concentrations were constant with storage at -80°C for ≥28 d. The Accu-Chek glucometer is an adequate instrument to measure honeybee glucose concentration in hemolymph diluted with 0.9% NaCl, with good accuracy and precision at <20 mmol/L. Hemolymph storage at -80°C is suitable for long-term conservation of glucose.

Citing Articles

Special section on honey bee health and disease.

Benoit-Biancamano M J Vet Diagn Invest. 2023; 35(6):595-596.

PMID: 37815222 PMC: 10621535. DOI: 10.1177/10406387231202959.

References
1.
Borsuk G, Ptaszynska A, Olszewski K, Domaciuk M, Krutmuang P, Paleolog J . A New Method for Quick and Easy Hemolymph Collection from Apidae Adults. PLoS One. 2017; 12(1):e0170487. PMC: 5268409. DOI: 10.1371/journal.pone.0170487. View

2.
Ryazanova A, Alekseev A, Slepneva I . The phenylthiourea is a competitive inhibitor of the enzymatic oxidation of DOPA by phenoloxidase. J Enzyme Inhib Med Chem. 2011; 27(1):78-83. DOI: 10.3109/14756366.2011.576010. View

3.
Chan Q, Howes C, Foster L . Quantitative comparison of caste differences in honeybee hemolymph. Mol Cell Proteomics. 2006; 5(12):2252-62. DOI: 10.1074/mcp.M600197-MCP200. View

4.
Tang Z, Du X, Louie R, Kost G . Effects of pH on glucose measurements with handheld glucose meters and a portable glucose analyzer for point-of-care testing. Arch Pathol Lab Med. 2000; 124(4):577-82. DOI: 10.5858/2000-124-0577-EOPOGM. View

5.
Buckemuller C, Siehler O, Gobel J, Zeumer R, Olschlager A, Eisenhardt D . Octopamine Underlies the Counter-Regulatory Response to a Glucose Deficit in Honeybees (). Front Syst Neurosci. 2017; 11:63. PMC: 5582081. DOI: 10.3389/fnsys.2017.00063. View