» Articles » PMID: 35976574

Locking in Trochanteric Fractures: a Comparison of Static Versus Dynamic Locking Using the Gamma3 Nail

Overview
Specialty Orthopedics
Date 2022 Aug 17
PMID 35976574
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: The Gamma3 nail (Stryker®) is an intramedullary device consisting of a proximal lag screw and distal interlocking screw. It is still unknown whether the screw locking mode could influence clinical outcomes. The aim of this study is to compare static and dynamic screw locking regarding their influence on surgical revisions and lag screw displacement.

Methods: A retrospective single-centre study was performed on patients ≥ 60 years admitted for a trochanteric fracture between September 2016 and January 2020. Surgical revisions and lag screw displacement were evaluated at 6 weeks and 1-year follow-up, respectively.

Results: A total of 142 patients were included for analysis. Surgical revisions were needed in 13 cases (9.2%). Indications included implant breakage (n = 3), lag screw cut-out (n = 3), lateral hip pain (n = 6) and non-union (n = 1). The number of surgical revisions was not different between static and dynamic locking (OR 2.55; 95%CI 0.73-8.56; p 0.142). The median lag screw displacement was 2.5 mm, which was similar for static and dynamic locking (2.3 mm versus 2.7 mm; p 0.785).

Conclusion: The screw locking mode of the Gamma3 nail is not associated with a higher risk of surgical revisions. However, the design of the Gamma3 nail may not be suitable for static locking.

Citing Articles

[Research progress on distal interlocking screws of cephalomedullary nails in intertrochanteric fractures].

Qi Y, Wang X, Chang S Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2024; 38(3):356-362.

PMID: 38500431 PMC: 10982033. DOI: 10.7507/1002-1892.202312052.


Development of early complications after treatment of trochanteric fractures with an intramedullary sliding hip screw in a geriatric population.

Robioneck M, Pishnamaz M, Becker N, Bolierakis E, Hildebrand F, Horst K Eur J Trauma Emerg Surg. 2023; 50(2):329-337.

PMID: 38081966 DOI: 10.1007/s00068-023-02404-8.


Biomechanical Behavior of Dynamic vs. Static Distal Locking Intramedullary Nails in Subtrochanteric Femur Fractures.

Martinez-Aznar C, Mateo J, Ibarz E, Gracia L, Rosell J, Puertolas S Bioengineering (Basel). 2023; 10(10).

PMID: 37892909 PMC: 10604699. DOI: 10.3390/bioengineering10101179.

References
1.
Sun D, Wang C, Chen Y, Liu X, Zhao P, Zhang H . A meta-analysis comparing intramedullary with extramedullary fixations for unstable femoral intertrochanteric fractures. Medicine (Baltimore). 2019; 98(37):e17010. PMC: 6750238. DOI: 10.1097/MD.0000000000017010. View

2.
Pascarella R, Fantasia R, Maresca A, Bettuzzi C, Amendola L, Violini S . How evolution of the nailing system improves results and reduces orthopedic complications: more than 2000 cases of trochanteric fractures treated with the Gamma Nail System. Musculoskelet Surg. 2015; 100(1):1-8. DOI: 10.1007/s12306-015-0391-y. View

3.
Chen M, Liao Y, Liu J, Fang W, Hong N, Ye X . Comparison of Sexual Knowledge, Attitude, and Behavior between Female Chinese College Students from Urban Areas and Rural Areas: A Hidden Challenge for HIV/AIDS Control in China. Biomed Res Int. 2017; 2016:8175921. PMC: 5215479. DOI: 10.1155/2016/8175921. View

4.
Yoo J, Kim T, Chang J, Kwak Y, Kwon Y . Factors influencing functional outcomes in united intertrochanteric hip fractures: a negative effect of lag screw sliding. Orthopedics. 2014; 37(12):e1101-7. DOI: 10.3928/01477447-20141124-58. View

5.
Parry J, Sapp T, Langford J, Koval K, Haidukewych G . Variables Associated With Lag Screw Sliding After Single-Screw Cephalomedullary Nail Fixation of Intertrochanteric Fractures. J Orthop Trauma. 2020; 34(7):356-358. DOI: 10.1097/BOT.0000000000001730. View