» Articles » PMID: 35928600

Technical Feasibility of Online Adaptive Stereotactic Treatments in the Abdomen on a Robotic Radiosurgery System

Overview
Specialty Oncology
Date 2022 Aug 5
PMID 35928600
Authors
Affiliations
Soon will be listed here.
Abstract

Background And Purpose: Stereotactic body radiotherapy (SBRT) has been proven to be beneficial for several disease sites in the (lower) abdomen. However, the quality of the treatment plan, based on a single planning computed tomography (CT), can be compromised due to large inter-fraction motion of the target and organs at risk (OARs) in this anatomical region. The aim of this study was to investigate the feasibility of online adaptive SBRT treatments on a robotic radiosurgery system and to record estimated total treatment times.

Materials And Methods: For two disease sites, locally advanced pancreatic cancer (LAPC) and oligometastatic lymph nodes, four patients with repeat CTs were included in the feasibility study. Quick treatment plan templates were generated based on the planning CT and validated by running them on the plan and fraction CTs. For two cases a dummy run was performed and the individual steps were timed. Dose delivery was the largest contributor to the total treatment time, followed by contour adaptation.

Results: Running the quick plan templates resulted in plans similar to unrestricted plans, obeying the OAR constraints. The dummy runs showed that online adaptive treatments were completed in 64 to 83 min respectively for oligometastases and LAPC, comparable to other clinically available solutions.

Conclusions: This study showed the feasibility of online re-planning for two challenging disease sites within a clinically acceptable time frame on a robotic radiosurgery system, making use of commercially available elements that are not integrated by the vendor.

Citing Articles

First-in-Men Online Adaptive Robotic Stereotactic Body Radiation Therapy: Toward Ultrahypofractionation for High-Risk Prostate Cancer Patients.

Brand V, Milder M, Christianen M, de Vries K, Hoogeman M, Incrocci L Adv Radiat Oncol. 2025; 10(2):101701.

PMID: 39866592 PMC: 11758839. DOI: 10.1016/j.adro.2024.101701.


Precision requirements in stereotactic arrhythmia radioablation for ventricular tachycardia.

Fast M, Lydiard S, Boda-Heggemann J, Tanadini-Lang S, Muren L, Clark C Phys Imaging Radiat Oncol. 2023; 28:100508.

PMID: 38026083 PMC: 10679852. DOI: 10.1016/j.phro.2023.100508.

References
1.
Ten Haken R, Forman J, Heimburger D, Gerhardsson A, McShan D, Schoeppel S . Treatment planning issues related to prostate movement in response to differential filling of the rectum and bladder. Int J Radiat Oncol Biol Phys. 1991; 20(6):1317-24. DOI: 10.1016/0360-3016(91)90244-x. View

2.
Lee M, Simeonov A, Stanescu T, Dawson L, Brock K, Velec M . MRI evaluation of normal tissue deformation and breathing motion under an abdominal compression device. J Appl Clin Med Phys. 2021; 22(2):90-97. PMC: 7882116. DOI: 10.1002/acm2.13165. View

3.
Kontaxis C, Bol G, Lagendijk J, Raaymakers B . A new methodology for inter- and intrafraction plan adaptation for the MR-linac. Phys Med Biol. 2015; 60(19):7485-97. DOI: 10.1088/0031-9155/60/19/7485. View

4.
Loi M, Magallon-Baro A, Suker M, van Eijck C, Sharma A, Hoogeman M . Pancreatic cancer treated with SBRT: Effect of anatomical interfraction variations on dose to organs at risk. Radiother Oncol. 2019; 134:67-73. DOI: 10.1016/j.radonc.2019.01.020. View

5.
Willigenburg T, de Muinck Keizer D, Peters M, Claes A, Lagendijk J, de Boer H . Evaluation of daily online contour adaptation by radiation therapists for prostate cancer treatment on an MRI-guided linear accelerator. Clin Transl Radiat Oncol. 2021; 27:50-56. PMC: 7822780. DOI: 10.1016/j.ctro.2021.01.002. View