Isocyanates May Contribute to Allergic Contact Dermatitis from Diabetes Devices and Wound Dressings

Overview
Publisher Wiley
Specialty Dermatology
Date 2022 Jul 11
PMID 35815495
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Isocyanates are well-known occupational allergens, but can also be present in medical devices.

Objectives: To highlight that contact sensitization to isocyanates might contribute to allergic contact dermatitis (ACD) from polyurethane (PU)-containing diabetes devices and wound dressings.

Patients And Methods: Nineteen patients with suspected ACD from diabetes devices and/or wound dressings were patch tested to an isocyanate series. Four wound dressings, six diabetes devices and four monomeric isocyanate patch test preparations were analysed with gas chromatography - mass spectrometry.

Results: Eight patients reacted to isocyanates and corresponding amines: 3 to isophorone diisocyanate (IPDI), 4 to 4,4'-diaminodiphenylmethane (MDA), 4 to 2,4-toluene diisocyanate (TDI) and 1 to polymeric methylene diphenyl diisocyanate (PMDI). Three of four wound dressings contained isocyanates (methylene diphenyl diisocyanate [MDI], TDI and/or IPDI), whereas five of six diabetes devices contained 4,4'-MDI, and one of them also IPDI. None of the medical devices contained 1,6-hexamethylene diisocyanate. Contrary to IPDI, and especially MDI, only the concentration of the TDI patch test preparation corresponded approximately (80%) to its label.

Conclusion: Patch tests with isocyanates may be worth-while in patients with suspected ACD from PU-containing medical devices. Besides MDA, and PMDI, also TDI might potentially be a marker for MDI-sensitization.

Citing Articles

Allergic Contact Dermatitis Induced by Modern Wound Dressings: A Comprehensive Analysis of Risks and Allergenic Components.

Kucuk K, Van Gysel J, Del Marmol V, White J Int Wound J. 2025; 22(3):e70153.

PMID: 40069101 PMC: 11896758. DOI: 10.1111/iwj.70153.


Allergic Contact Dermatitis Following Hypoglossal Nerve Stimulator Implant.

Tangutur A, Brown K, Dedhia R JAMA Otolaryngol Head Neck Surg. 2025; .

PMID: 40014338 PMC: 11869092. DOI: 10.1001/jamaoto.2024.5446.


Allergic contact dermatitis caused by glucose sensors and insulin pumps: A full review: Part 2. Case reports and case series, clinical features, patch test procedures, differentiation from irritant dermatitis, management of allergic patients and....

de Groot A, van Oers E, Ipenburg N, Rustemeyer T Contact Dermatitis. 2024; 92(3):164-175.

PMID: 39600134 PMC: 11795346. DOI: 10.1111/cod.14697.


Allergic contact dermatitis caused by glucose sensors and insulin pumps: A full review: Part 1: Sensors and pumps, adverse cutaneous reactions, allergens, and diabetes devices causing allergic contact dermatitis.

de Groot A, van Oers E, Ipenburg N, Rustemeyer T Contact Dermatitis. 2024; 92(2):87-112.

PMID: 39600106 PMC: 11710928. DOI: 10.1111/cod.14698.


Insulin Pump Therapy and Adverse Skin Reactions With Focus on Allergic Contact Dermatitis in Individuals Living With Diabetes Mellitus: A Systematic Review and Clinical-Based Update.

von Kobyletzki L, Ulriksdotter J, von Kobyletzki E, Mowitz M, Jendle J, Svedman C J Diabetes Sci Technol. 2024; 18(6):1300-1312.

PMID: 38853748 PMC: 11535278. DOI: 10.1177/19322968241252613.