Comparison of Anterior Segment Measurements with a New Multifunctional Unit and Five Other Devices

Overview
Specialty Ophthalmology
Date 2022 Jun 29
PMID 35766048
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To evaluate the clinical availability of a multifunctional ocular biometric unit, MR-6000, for simultaneous keratometry, tonometry, topography, and pachymetry evaluation, and compare anterior segment measurements with five other devices: autokeratometer (KR-1), Scheimpflug camera (Pentacam HR), swept-source optical coherence tomography (IOLMaster 700), Placido disk scanning-slit topography (Orbscan II), and noncontact tonometry (FT-1000).

Methods: Thirty eyes from thirty patients who visited Severance Hospital for cataract surgery were examined using MR6000 and the other devices. The mean keratometry, central corneal thickness (CCT), white-to-white (WTW) distance, and intraocular pressure (IOP) values were compared. Repeated measures analysis of variance, Wilcoxon signed-rank test, intraclass correlation coefficient (ICC), and Bland-Altman plot were used to assess the correlation and agreement between devices.

Results: Thirty eyes of thirty patients were evaluated. Statistically significant differences in mean keratometry between MR6000, KR-1, Pentacam HR, and IOLMaster 700 were not observed (p > 0.05). All five devices, including Orbscan II, had almost perfect agreement in measuring keratometry (ICC > 0.80, p < 0.05). CCT measured by MR-6000 was significantly different from that of Pentacam HR and Orbscan II measurements (p < 0.05) but correlated with that of Pentacam HR and Orbscan II measurements (ICC > 0.60, p < 0.05). The WTW distance measured by MR-6000 was not significantly different from that measured by IOLMaster 700 but was different from that measured by Orbscan II. IOP measured by MR-6000 was not correlated with FT-1000.

Conclusions: Keratometric values obtained through MR-6000 can be used interchangeably with other devices based on good correlation and agreement. However, the CCT, WTW, and IOP values were not interchangeable with a single multifunctional unit for cataract surgery preoperative examination.

References
1.
Meyer J, Gokul A, Vellara H, Prime Z, McGhee C . Repeatability and Agreement of Orbscan II, Pentacam HR, and Galilei Tomography Systems in Corneas With Keratoconus. Am J Ophthalmol. 2016; 175:122-128. DOI: 10.1016/j.ajo.2016.12.003. View

2.
Oh J, Shin H, Kim J, Kim H, Song J . Direct measurement of the ciliary sulcus diameter by 35-megahertz ultrasound biomicroscopy. Ophthalmology. 2007; 114(9):1685-8. DOI: 10.1016/j.ophtha.2006.12.018. View

3.
Graf M . [Significance of the corneal thickness in non-contact tonometry]. Klin Monbl Augenheilkd. 1991; 199(3):183-6. DOI: 10.1055/s-2008-1046068. View

4.
Reinstein D, Lovisolo C, Archer T, Gobbe M . Comparison of postoperative vault height predictability using white-to-white or sulcus diameter-based sizing for the visian implantable collamer lens. J Refract Surg. 2013; 29(1):30-5. DOI: 10.3928/1081597X-20121210-02. View

5.
Hoffer K, Hoffmann P, Savini G . Comparison of a new optical biometer using swept-source optical coherence tomography and a biometer using optical low-coherence reflectometry. J Cataract Refract Surg. 2016; 42(8):1165-72. DOI: 10.1016/j.jcrs.2016.07.013. View