Accuracy of Acetabular Cup Placement During Total Hip Arthroplasty in Supine Position Using a Pelvic Rotation Correction Device

Overview
Publisher Elsevier
Date 2022 Jun 1
PMID 35647246
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Accurate cup placement during total hip arthroplasty (THA) is difficult because the intraoperative pelvic position changes even in supine patient position. We developed a device known as HipPointer; it corrects pelvic rotation and creates a functional pelvic plane as a reference. The aim of this study was to determine the device placement accuracy and investigate causes of error.

Material And Methods: HipPointer was used for cup placement in 353 hips of 308 patients who underwent direct-anterior-approach THA in supine position. The mean age at surgery and body mass index were 63.9 (17-90) years and 24.9 (16.6-42.0) kg/m, respectively. The mean observation period was 40.5 (12-73) months. To investigate the accuracy of HipPointer, preoperative planning and postoperative cup placement angles relative to the functional pelvic plane were evaluated using a three-dimensional analysis software, and absolute errors were determined.

Results: The means ± standard deviations of radiographic inclination (RI) and radiographic anteversion (RA) were 40.2 ± 3.0° and 15.8 ± 3.6°, respectively. The absolute errors of RI and RA were 2.2 ± 2.0° and 2.7 ± 2.3°, respectively. The ratio of the cup placement angle for which both RI and RA are ≤10° in the target zone was 99% (350/353 hips), and the ratio of the absolute errors for which both RI and RA are ≤5° was 80.4% (284/353 hips).

Conclusions: HipPointer is simple in structure, easy to use, and useful for direct-anterior-approach THA in supine position. It provides good cup placement accuracy.

Citing Articles

Secure paediatric pelvic fracture external fixation using an intraoperative support device.

Igei T, Nakasone S, Onaga M, Nishida K BMJ Case Rep. 2023; 16(4).

PMID: 37011998 PMC: 10083815. DOI: 10.1136/bcr-2022-252365.

References
1.
Nishihara S, Sugano N, Nishii T, Ohzono K, Yoshikawa H . Measurements of pelvic flexion angle using three-dimensional computed tomography. Clin Orthop Relat Res. 2003; (411):140-51. DOI: 10.1097/01.blo.0000069891.31220.fd. View

2.
Ando W, Takao M, Hamada H, Uemura K, Sugano N . Comparison of the accuracy of the cup position and orientation in total hip arthroplasty for osteoarthritis secondary to developmental dysplasia of the hip between the Mako robotic arm-assisted system and computed tomography-based navigation. Int Orthop. 2021; 45(7):1719-1725. DOI: 10.1007/s00264-021-05015-3. View

3.
Barrett W, Turner S, Leopold J . Prospective randomized study of direct anterior vs postero-lateral approach for total hip arthroplasty. J Arthroplasty. 2013; 28(9):1634-8. DOI: 10.1016/j.arth.2013.01.034. View

4.
Sugano N, Takao M, Sakai T, Nishii T, Miki H . Does CT-based navigation improve the long-term survival in ceramic-on-ceramic THA?. Clin Orthop Relat Res. 2012; 470(11):3054-9. PMC: 3462880. DOI: 10.1007/s11999-012-2378-4. View

5.
Hayashi S, Hashimoto S, Kuroda Y, Nakano N, Matsumoto T, Ishida K . Accuracy of cup position following robot-assisted total hip arthroplasty may be associated with surgical approach and pelvic tilt. Sci Rep. 2021; 11(1):7578. PMC: 8027591. DOI: 10.1038/s41598-021-86849-0. View