Abstract

Polymerase chain reaction (PCR) is limited by the long reaction time for point-of-care. Currently, commercial benchtop rapid PCR requires 30−40 min, and this time is limited by the absence of rapid and stable heating and cooling platforms rather than the biochemical reaction kinetics. This study develops an ultrafast PCR (<3 min) platform using flow-through microchannel chips. An actin gene amplicon with a length of 151 base-pairs in the whole genome was used to verify the ultrafast PCR microfluidic chip. The results demonstrated that the channel of 56 μm height can provide fast heat conduction and the channel length should not be short. Under certain denaturation and annealing/extension times, a short channel design will cause the sample to drive slowly in the microchannel with insufficient pressure in the channel, causing the fluid to generate bubbles in the high-temperature zone and subsequently destabilizing the flow. The chips used in the experiment can complete 40 thermal cycles within 160 s through a design with the 56 µm channel height and with each thermal circle measuring 4 cm long. The calculation shows that the DNA extension speed is ~60 base-pairs/s, which is consistent with the theoretical speed of the Klen Taq extension used, and the detection limit can reach 67 copies. The heat transfer time of the reagent on this platform is very short. The simple chip design and fabrication are suitable for the development of commercial ultrafast PCR chips.

Citing Articles

Liquid biopsy: Comprehensive overview of circulating tumor DNA (Review).

Ge Q, Zhang Z, Li S, Ma J, Zhao Z Oncol Lett. 2024; 28(5):548.

PMID: 39319213 PMC: 11420644. DOI: 10.3892/ol.2024.14681.


Customizable Nichrome Wire Heaters for Molecular Diagnostic Applications.

Lim J, Han W, Thang L, Lee Y, Shin J Biosensors (Basel). 2024; 14(3).

PMID: 38534259 PMC: 10968288. DOI: 10.3390/bios14030152.


Assessing Different PCR Master Mixes for Ultrarapid DNA Amplification: Important Analytical Parameters.

Brukner I, Paliouras M, Trifiro M, Bohbot M, Shamir D, Kirk A Diagnostics (Basel). 2024; 14(5).

PMID: 38472949 PMC: 10931018. DOI: 10.3390/diagnostics14050477.


LoCKAmp: lab-on-PCB technology for <3 minute virus genetic detection.

Papamatthaiou S, Boxall-Clasby J, Douglas E, Jajesniak P, Peyret H, Mercer-Chalmers J Lab Chip. 2023; 23(20):4400-4412.

PMID: 37740394 PMC: 10563828. DOI: 10.1039/d3lc00441d.


Advances in Simple, Rapid, and Contamination-Free Instantaneous Nucleic Acid Devices for Pathogen Detection.

Wang Y, Wang C, Zhou Z, Si J, Li S, Zeng Y Biosensors (Basel). 2023; 13(7).

PMID: 37504131 PMC: 10377012. DOI: 10.3390/bios13070732.


References
1.
Liu W, Zhang M, Liu X, Sharma A, Ding X . A Point-of-Need infrared mediated PCR platform with compatible lateral flow strip for HPV detection. Biosens Bioelectron. 2017; 96:213-219. DOI: 10.1016/j.bios.2017.04.047. View

2.
Millington A, Houskeeper J, Quackenbush J, Trauba J, Wittwer C . The kinetic requirements of extreme qPCR. Biomol Detect Quantif. 2019; 17:100081. PMC: 6591793. DOI: 10.1016/j.bdq.2019.100081. View

3.
Sun Y, Kwok Y, Foo-Peng Lee P, Nguyen N . Rapid amplification of genetically modified organisms using a circular ferrofluid-driven PCR microchip. Anal Bioanal Chem. 2009; 394(5):1505-8. DOI: 10.1007/s00216-009-2808-7. View

4.
Kuo Y, Li Y, Chan E . Rapid identification of HPV 16 and 18 by multiplex nested PCR-immunochromatographic test. J Virol Methods. 2014; 212:8-11. DOI: 10.1016/j.jviromet.2014.10.009. View

5.
Wu W, Kang K, Lee N . Bubble-free on-chip continuous-flow polymerase chain reaction: concept and application. Analyst. 2011; 136(11):2287-93. DOI: 10.1039/c0an01034k. View